Page 5.2
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Chapter 5
Adjustments
Introduction
This Chapter provides the procedures necessary to check and adjust all service settings. Every
procedure is self-contained and details the tools required to carry out adjustments, and the test
equipment necessary. Any adjustments, jumper settings, special checks or functional tests that
are required on the module, or on associated modules or software, are also detailed.
In-text diagrams support the text where required and provide location details of connectors, test
points and adjustment potentiometers.
Specific warnings and cautions are given in the text where applicable.
The part numbers for all replaceable modules are given in Chapter 6.
The AT-2plus has the following adjustments:
•
Battery Charge Voltage (VR1 on the main board MK 14-10)
•
ECG Amplifier reference voltage and ramp slope (VR2 and VR3 on the main board MK14-
10)
We recommend that the reference voltages are checked every year.