Page 4.16
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Chapter 4
Module Removal and Replacement
RS-232 Serial Interface Board MK 14 - 11
Prerequisite
•
The Warnings and Cautions at the beginning of the Chapter must be observed.
•
The Top Assembly must be removed as detailed previously. All external cable assemblies
must be disconnected.
Tools
•
Cross-bladed screwdriver
Part Number
The part number for the RS-232 board is given in Chapter 6.
CAUTION
THE RS-232 BOARD CONTENT IS STATIC SENSITIVE; OBSERVE ANTISTATIC
PRECAUTIONS
RS-232 board Removal
1. Disconnect the 10 pole connector to the main board.
2. Unscrew the four retaining screws securing the board in position.
3. Remove the board.
RS-232 board Replacement
To replace the RS-232 board proceed as follows:
1. Position the board and secure at the 4 fixing points.
2. Connect the 10 pole connector to the main board.
3. Reassemble the unit.