Page 1.28
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Chapter 1
Operating Elements
Software Updating via RS-232
The AT-2
plus Memory
software can be updated by using a PC program called "SWUP".
Note !
This is only possible with the AT-2plus Memory with software version higher than V
2.11 ! The SWUP programme needs to be reinstalled for each new software.
Prerequisites
•
RS-232 cable assembly P/N 2.310 159
•
SWUP Programme for Win 95 ( 2 disks or Zip-file from Mailbox)
Procedure
1. Connect the AT-2plus Memory with the RS-232 cable to any free COM port on your PC
2. Switch the PC ON and go to Win 95.
3. Install the SWUP programme by using "Install.exe" on diskette No.1.
4. Follow the instructions given by the programme.
5. Access the programme through the icon "SWUP". (Schiller-Menu)
6. Go to the configuration menu and chose the applicable COM port (1 or 2).
7. Set the Baudrate to "Auto".
8. Make sure that the AT-2plus is switched ON.
9. Chose "Start 0", "Start 1" or "Start 2".
10. "Start 0" loads the software with languages GER, ENG, FRE, SWE and USA in the AT-2plus
Memory.
11. "Start 1" loads the software with languages ITA, SPA, and POR in the AT-2plus Memory.
12. "Start 2" loads the software with language RUS in the AT-2plus Memory (from version V
2.20). Russian interpretation must be set by setting Russian language with ALT - 0 - 2 - 9.