3. TECHNICAL SPECIFICATIONS
Flow
Rate
Micro
:
0.02 – 4.00 ml/min
Analytical
:
0.05 – 9.95 ml/min
Semi
-
Preparative
: 0.20 – 40.0 ml/min
Pump
Head
Material
:
Stainless Steel, PEEK, or Titanium
Pulsation
:
less than 1.0 %
Flow Accuracy
:
less than 0.1 %
Flow Precision
:
less than 1.0 %
Max. Pressure
40 MPa (400 bar)
Display
:
Alphanumeric, 2x20 characters LCD
Compressibility Factor:
selectable from 0.7 to 1.0
Operation Mode:
constant flow or constant pressure
Remote Control Analog:
Flow Rate (0-10V)
;
Start, Stop
Digital:
RS232
Programmable Runtime:
00:01 to 99:59 h
Programmable Delay Time:
00:01 to 99:59 h
Safety Control:
Min./Max. Pressure
Power Supply
:
220/110 Volt; 50/60 Hz
Dimensions
:
258 x 148 x 350 mm
Weight
:
8 kg