SDS 9414i HPLC Pump
Version: 2.2 (July 2007)
Page: 41 / 45
In case of a wrong checksum, the pump will answer with a “
?
”. Otherwise, when the command is
correct and understood by the pump it answers with a string:
Byte 1
Length
This states the total length of the answer, including
the checksum byte and its own length. Normally, the
answer has a length of 4 bytes.
Byte 2
Pump Status
This states the pump’s status:
Bit
Value
7
Remote start accepted (1 = TRUE)
6
Not used
5
Pressure Failure (1 = TRUE)
4
Not used
3
Not used
2
Pump Head mounted (1 = TRUE)
1
Preparative Pump Head (1 = TRUE)
0
Micro Pump Head (1 = TRUE)
When Bit 0 and 1 are both “0”, a Analytical Pump
Head is mounted.
Byte 3
Pressure
This byte states the current pump pressure. Actual
value in MPa is the received byte x 0.2.
Byte 4
Checksum
This checksum is calculated as described above.
The answer is always terminated with a “
.
”.