SDS 9414i HPLC Pump
Version: 2.2 (July 2007)
Page: 19 / 45
The HPLC pump SDS 9414I has a built-in start delay time function which protects and
therefore prolongs column's durability as a careful increase of pressure onto the column is
possible.
During the start delay time operation, the display shows the symbol “
^
” right before the
RT
.
The start delay time can be selected for a time range of up to 2 min 30 sec at a maximum.
The time value is selected by using the [
p
] or [
q
] key. The [
PROG
] key has to be pushed to
fix the value into the memory and to activate the next programming step.
5.9 STOP DELAY TIME
By pushing the [
PROG
] key the programming dialog is switched to the next menu. The display
shows:
The HPLC pump SDS 9414I has a built-in stop delay time function which protects and
therefore prolongs column's durability as a careful decrease of pressure onto the column is
possible.
During the stop delay time operation, the display shows the symbol “
v
” right before the
RT
.
The stop delay time can be selected for a time range of up to 2 min 30 sec at a maximum.
The time value is selected by using the [
p
] or [
q
] key. The [
PROG
] key has to be pushed to
fix the value into the memory and to leave the PROGRAMMING MODE.