SDS 9414i HPLC Pump
Version: 2.2 (July 2007)
Page: 16 / 45
By pushing the [
PROG
] key this value will be fixed into the memory and the next program step is
called up.
5.2.2 MAXIMUM FLOW
By pushing the [
PROG
] key the programming dialog is switched to the next menu. The display
shows:
This step is built in as a safety feature to avoid any damage of the analyzer by reaching a too
high flow rate.
The desired maximum flow rate can be set by using the [
p
] or [
q
] key.
5.3 MINIMUM PRESSURE THRESHOLD
By pushing the [
PROG
] key the programming dialog is switched to the next menu. The display
shows:
At this step the minimum pressure level can be set. Any value from 0.0, but no higher than the
maximum level. A practical level is a value of about 20 % of the normal operation pressure.
After setting the wanted level by using the [
p
] or [
q
] key, the [
PROG
] key has to be used to
fix this value in the memory and to activate the next program step.
5.4 MAXIMUM PRESSURE THRESHOLD
By pushing the [
PROG
] key the programming dialog is switched to the next menu. The display
shows: