56
SBS-3510 User’s Manual
5. Reassemble all parts in reverse order.
14.1.2 Exchanging the Pump Lever
1. Dismount the pump as described in section 12.3.
2. The pump lever sits on the two pivot pins. Pull it off carefully by hand.
Do not use any tools.
3. Mount the new pump lever by shifting it in place.
4. Remount the pump as described in section 12.3.
14.2 Software Update
Your SBS representative will inform you when a new software update for
your SBS-3510 is available. After you have received the update file, you can
import it into the instrument via Bluetooth.
IMPORTANT:
Ensure that the batteries of the instrument are fully charged
before you start a software update.
NOTICE
• Observe the correct polarity when you insert the batteries, see
the engraving in the battery compartment.
• Use only batteries of the same type and with the same charge level.
• Ensure that the interior of the instrument is perfectly dry. Any kind
of liquid will cause damage to the electronic parts of the instrument.
WARNING
The cover of the battery compartment must always be closed when you
use the instrument.
NOTICE
All data and settings will be deleted or reset to factory settings during the
update.
• Export and save all important data (methods, sample IDs, custom
parameters, measured data, etc.) before you perform a software update.