SBS-3510 User’s Manual
8.3 Performing a Custom Adjustment
You can use any reference liquid with a known density at 20°C (68°F) and
a known temperature coefficient Alpha to perform a custom adjustment.
We recommend to perform a custom adjustment if, for example, you measure
only liquids in a specific density range (e.g. high density) and want to achieve
more accurate results in this range.
1. Clean the measuring cell as described in section 12.1.
2. Rinse the measuring cell until no residues of cleaning agent are present
anymore.
3. Tap <Menu> and select
Adjustments > Custom Adjustment
.
4. Fill in your reference liquid with 15°C to 25°C (59°F to 77°F).
5. Tap <Start>.
6. Tap <Edit> and enter the density of the reference liquid at 20°C (68°F).
The density value must be in the range 0.5–1.5 g/cm
3
.
7. Tap <Next>.
8. Tap <Edit> and enter the temperature coefficient Alpha of the reference
liquid.
The value for the temperature coefficient must be in the range 0–0.00999.
9. Tap <Next>.
10. After the procedure has finished, the density deviation from the previous
adjustment is displayed.
11. To save the new adjustment, tap <Save>.
To reject the adjustment, tap <Cancel>.
8.4 Reset to Factory Adjustment
If you want to undo all your adjustments, reset the instrument to factory
adjustment.
1. Tap <Menu> and select
Adjustments > Factory Adjustment
.
2. Tap <Yes> to reset the instrument to factory adjustment.
The factory adjustment will be restored.
3. Tap <Back> repeatedly to return to the main screen.