![SBS SBS-3510 Instruction Manual And Safety Information Download Page 30](http://html1.mh-extra.com/html/sbs/sbs-3510/sbs-3510_instruction-manual-and-safety-information_1205394030.webp)
30
SBS-3510 User’s Manual
Deleting methods
1. Tap <Menu> and select
Methods > Edit Method
.
2. Select a method that you want to delete, and tap <Delete>.
You cannot delete the last remaining method.
3. Tap <Yes> to confirm deletion.
Tap <No> to abort deletion.
4. Tap <Back> repeatedly to return to the main screen.
7.2.2 Assigning a Method to a Measurement
1. In the quick access area, activate
(“Sample” function).
2. Select “Method” and tap <Edit>.
3. Select the appropriate method from the list.
4. Tap <Back> to return to the main screen.
All subsequent measurements will use the selected method until you assign
a new one.
The assigned method is shown in the header.
7.2.3 Defining Limits
The function will give a warning when measurement results of parameter 1
lie outside the margins of your internal quality specification.
Define limits in the method settings. Measurement results outside the thereby
defined margins will be highlighted by a yellow background color and marked
in the data memory accordingly.
1. Tap <Menu> and select
Methods > Edit Method
.
2. Select a method from the list and tap <Edit>.
3. Select “Limit” and tap <Yes> to activate the limit function.
Two more input fields for entering the limit values come up in the
parameter list.
4. Enter the values for “Upper Limit” and “Lower Limit”.
IMPORTANT:
The limits apply only to parameter 1.
5. Tap <Back> repeatedly to return to the main screen.