11. 

Notes 

 
With  the  LN  function,  the  linearization  of  the  measuring  instrument  specified  by  the 
calibration  can  be  changed.  However,  it  is  strongly  recommended  not  to  make  any 
changes to the 

LN value, as

 this change will lead to deviating measurement results. 

Any change in the Ln value can significantly affect the measuring accuracy. This 
value should only be set by qualified personnel. 

 In general, the larger the Ln value, the smaller the reading result for the same layer 
thickness. A small change of the Ln- value causes a large       change in the reading 
result in the upper measuring range (at 500µm/ 20mil). 
Thus the Ln value is to be corrected: 
Pressing the on/off button: It takes about 11 seconds from the beginning of pressing 
the on/off button. 
Its value can be changed by pressing the Plus or Minus key after 'LN' appears in the 
display  and  releasing  the  On/Off  key.  The  value  is  stored  and  then  the  zero  key  is 
pressed.     
A: The reading result in the lower area is corrected by pressing the Plus or Minus key. 
B:  The Ln  value  is  increased  if  the  reading  result  is  correct  in  the  lower  range  (e.g. 
51µm) but too large in the upper range (e.g. 432µm). 
In contrast, the Ln value is reduced if the reading result is correct in the lower range 
(e.g. 51µm) but too small in the upper range (e.g. 432µm).    
C: This procedure of A. and B. is repeated until the reading result is satisfactory in its 
accuracy for each distance foil. 
 
Note:  
To view the CE declaration, please click on the following link:  

https://www.kern-sohn.com/shop/de/DOWNLOADS/

 

Summary of Contents for TC 1250-0.1F

Page 1: ...2336 Balingen e mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version 2 0 04 2020 GB TC BA e 2020 PROFESSIONAL MEASURING ...

Page 2: ...lity of this measuring device and its extensive functionality For any questions wishes and suggestions please do not hesitate to contact us Table of contents 1 Introduction 3 1 1 General description 3 2 Functions 3 3 Technical data 3 4 Description of the control panel 5 5 Measuring procedure 5 6 Zero adjustment 6 7 Battery replacement 6 8 Adjusting foils 7 9 General notes 7 10 Restore factory sett...

Page 3: ... etc This probe is often used to measure the thickness of galvanized layers lacquer layers enamelled porcelain layers phosphorescent layers copper plates aluminium plates alloys paper etc The N probe measures the thickness of non magnetic layers on non magnetic metals It is used to measure anodizations lacquer layers glazes colours enamel plastic layers powder coating etc These should be on non ma...

Page 4: ...ries UM 4 Ambient temperature 0 C to 50 C Air humidity 80 Dimensions 126 x 65 x 27mm Weight Ca 81g without batteries Scope of delivery Carrying case operating instructions Depending on the type 1 built in measuring probe F TC 1250 0 1 F with F N key 2 built in measuring probe N TC 1250 0 1 N with F N key 3 built in measuring probe FN TC 1250 0 1 FN without F N u S C key Spacer foils for each model...

Page 5: ...tself by the symbol Fe F for ferrous metals or NFe N for non ferrous metals shown on the display It switches to the automatic measuring mode which also correctly recognizes and assigns the zero plate or any other substrate itself b Place the measuring probe 4 1 on the layer to be measured The layer thickness can now be read on the display The reading result can be corrected by pressing the plus ke...

Page 6: ...ressing the on off button 6 Zero adjustment a The zero adjustment for Fe F and NFe N should be done separately The iron zero plate is taken when Fe is read on the display while the aluminium zero plate is taken when NFe is read The measuring probe 4 1 is now carefully placed on the zero plate The zero on off key 4 6 is pressed and 0 is shown in the display without lifting the probe Attention The z...

Page 7: ...probe should only be carried out by qualified personnel can be carried out 10 Restore factory settings a It is recommended that the factory settings be performed in the following cases no more measurements can be made at all the measuring accuracy is impaired due to drastic changes in the environment in which the measuring instrument is used After replacing the measuring probe b Procedure The fact...

Page 8: ...ng the on off button It takes about 11 seconds from the beginning of pressing the on off button Its value can be changed by pressing the Plus or Minus key after LN appears in the display and releasing the On Off key The value is stored and then the zero key is pressed A The reading result in the lower area is corrected by pressing the Plus or Minus key B The Ln value is increased if the reading re...

Reviews: