1. 

Introduction 

1.1  General description 

 
This coating thickness gauge is small, light and handy. Although it has complex and 
advanced equipment, it is convenient and easy to use. 
Its robustness allows it to be used for many years, provided that all instructions in this 
manual are carefully followed.  
Therefore please keep them always within reach! 

Note:  It  is  strongly  recommended  to  adjust  the  new meter  before  first  use,  as 
described in chapter 6. This will result in a higher measurement accuracy from 
the beginning. 

 

2. 

Functions 

 
"This device complies with the ISO- Norm 2360 as well as the DIN-Norm, ASTM and 
BS. This means that it can be used both under laboratory conditions and under rough 
environmental conditions "in the field". 
"The  F"  probe  measures  the  thickness  of  non-magnetic  layers,  e.g.  paint,  plastic, 
enamelled porcelain, copper, zinc, aluminium, chrome, lacquer layers etc.  
These  layers  should  be  on  magnetic  materials  such  as  steel,  iron,  nickel,  etc.  This 
probe  is  often  used  to  measure  the  thickness  of  galvanized  layers,  lacquer  layers, 
enamelled porcelain layers, phosphorescent layers, copper plates, aluminium plates, 
alloys, paper, etc. 
"The  N-  probe  measures  the  thickness  of  non-magnetic  layers  on  non-magnetic 
metals. 
It  is  used  to  measure  anodizations,  lacquer  layers,  glazes,  colours,  enamel,  plastic 
layers,  powder  coating  etc.  These  should  be  on  non-magnetic  substrates  such  as 
aluminium, sheet metal, non-magnetic stainless steel and others. 
"The FN measuring probe is a combination of both probes and is therefore defined for 
both areas of application 
"Automatic recognition of the carrier material 
"Manual or automatic switch-off to conserve battery power. 
"Metric/ imperic conversion of the measured values of µm/mil 
"Two measuring modes: - single and continuous (S= single/ C= continuous) 
"Wide measuring range and high resolution 
" Can be connected to a PC for data transfer via RS 232 interface 
 Cable and software are available as optional accessories (ATC-01). 

3. 

Technical data 

 

TC 1250-0.1F 

TC 1250-0.1FN  TC 1250-0.1N 

Display 

4 digits, 10mm LCD display with backlight 

Measuring range 

0 to 1250 

μm / 0 to 50 mil (standard) 

Summary of Contents for TC 1250-0.1F

Page 1: ...2336 Balingen e mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version 2 0 04 2020 GB TC BA e 2020 PROFESSIONAL MEASURING ...

Page 2: ...lity of this measuring device and its extensive functionality For any questions wishes and suggestions please do not hesitate to contact us Table of contents 1 Introduction 3 1 1 General description 3 2 Functions 3 3 Technical data 3 4 Description of the control panel 5 5 Measuring procedure 5 6 Zero adjustment 6 7 Battery replacement 6 8 Adjusting foils 7 9 General notes 7 10 Restore factory sett...

Page 3: ... etc This probe is often used to measure the thickness of galvanized layers lacquer layers enamelled porcelain layers phosphorescent layers copper plates aluminium plates alloys paper etc The N probe measures the thickness of non magnetic layers on non magnetic metals It is used to measure anodizations lacquer layers glazes colours enamel plastic layers powder coating etc These should be on non ma...

Page 4: ...ries UM 4 Ambient temperature 0 C to 50 C Air humidity 80 Dimensions 126 x 65 x 27mm Weight Ca 81g without batteries Scope of delivery Carrying case operating instructions Depending on the type 1 built in measuring probe F TC 1250 0 1 F with F N key 2 built in measuring probe N TC 1250 0 1 N with F N key 3 built in measuring probe FN TC 1250 0 1 FN without F N u S C key Spacer foils for each model...

Page 5: ...tself by the symbol Fe F for ferrous metals or NFe N for non ferrous metals shown on the display It switches to the automatic measuring mode which also correctly recognizes and assigns the zero plate or any other substrate itself b Place the measuring probe 4 1 on the layer to be measured The layer thickness can now be read on the display The reading result can be corrected by pressing the plus ke...

Page 6: ...ressing the on off button 6 Zero adjustment a The zero adjustment for Fe F and NFe N should be done separately The iron zero plate is taken when Fe is read on the display while the aluminium zero plate is taken when NFe is read The measuring probe 4 1 is now carefully placed on the zero plate The zero on off key 4 6 is pressed and 0 is shown in the display without lifting the probe Attention The z...

Page 7: ...probe should only be carried out by qualified personnel can be carried out 10 Restore factory settings a It is recommended that the factory settings be performed in the following cases no more measurements can be made at all the measuring accuracy is impaired due to drastic changes in the environment in which the measuring instrument is used After replacing the measuring probe b Procedure The fact...

Page 8: ...ng the on off button It takes about 11 seconds from the beginning of pressing the on off button Its value can be changed by pressing the Plus or Minus key after LN appears in the display and releasing the On Off key The value is stored and then the zero key is pressed A The reading result in the lower area is corrected by pressing the Plus or Minus key B The Ln value is increased if the reading re...