SW-BA-e-2020
39
If the user selects the
Manual
option, the user
must click
Print Now
and
press
the <Enter> key
to print the measurement data.
☆
Note : Switch to the
Printer
mode in the
RS-232
menu before you start printing.
4.7 Factory settings
Factory settings
offer the possibility to reset all parameters that
have been changed by the users to the factory settings. The
parameters are initialized with the default values. Use the arrow
keys <◄>, <►> to select Y (Yes) or
N
(No). If you select
Y
(Yes)
and
press
the <Enter> key, the parameter will be initialized. If you
select
N
(No) or press the
ESC key,
the reset will be cancelled.
5. data transmission protocol RS-232
The
SW 1000/2000
sound level meter is equipped with a serial interface RS-232 The
user can modify the configuration of the sound level meter via the serial interface and
start and stop the sound level meter, request the current measured value parameters
and process the results further. The operation via the serial interface does not change
the operation via the keyboard.
5.1 Hardware configuration and interface settings
The
SW 1000/2000
device uses a 3-wire serial interface, the physical socket is a
6-pin PS/2 socket. The following is the definition of the RS-232 interface:
Transmission settings for RS-232:
Transmission mode
Full duplex
Synchronous
/asynchronous
mode
Asynchronous transmission
Baud rate
4800 bps, 9600 bps, 19200 bps
Data bits
8 bits
Stop bits
1 bit
Parity
None
Data flow control
Observe the time data in the table with the
nominal parameters.
1. NC
2. GRO
UND
3. NC
4. TXD
5. NC
6. RXD
3
2
1
4
5
6