4.3.1 Measurement setup
The
MEAS.setup
menu is the most important menu for the
measurement. Here you can set the parameters for
delay
,
integration period
,
repeat
,
SW protocol step
,
SWN protocol
step
,
CSD protocol
and
CSD protocol step.
You can select
options with the arrow keys <▲>, <▼>.
Delay
(Delay:
The delay time between pressing the <Start> key
and
the start of the
measurement. With the arrow keys <◄>, <►> you can select the delay time as
follows Sync 1 min, Sync 15 min, Sync 30 min, Sync 1 h, 1 s to 60 s.
The delay time prevents the measurement from being influenced by the keystroke
or by vibrations.
Itg. period
(integral period):
Itg.Period
is the integral period for each individual measurement. At the beginning
of each integral period, all integral data and time data are reset; the indication of
an overload or underrange is deleted. Integral data and time data are LEQ, Max,
Min, Peak, SD, SEL, E and L
N. Press <◄>, <►> to select the following option:
Infinite, 1 s to 59 s, 1 min to 59 min, 1 h to 24 h.
Repeat
:
The number of repetitions in a measurement. Total measuring time =
integral
period
x
repetition
. Press <◄>, <►> to select the following option: Inf, 1 through
9999.
SWN Logger
(SWN Protocol:
Press the arrow keys <◄>
,<►>
to switch over. When you select this option, the
sound pressure meter saves the data in SWN/OCT files.
The SWN/OCT files store the time history data. The data source in
Sound
Pressure Gauge Mode
is Profile 1 - 3 (selected in the SWN Save option of the
Profile 1 - 3 menu); the data is saved as an SWN file. In 1/1 octave mode, all
octave bands as well as LAeq, LBeq, LCeq, LZeq are saved as an OCT file.
SWN Log Step
(SWN log step)
:
SWN Log Step
(SWN log step) is the log step (the interval time) for storing data
as SWN/OCD time. Press <◄>, <►> to select the following option: 0.1s, 0.2s,
0.5s, 1s to 59s, 1min to 59min, 1h to 24h CSD Logger
protocol
):
Press the arrow keys <◄>
,<►>
to switch over. When the appropriate option is
selected, the sound pressure meter saves the results in CSD files.
CSD-files save the current data. The data source in the
sound pressure meter
mode
are the 14 group results of the
adjustable measurement
; these are stored
as CSD files. In the 1/1 octave mode, all octave bands as well as LAeq, LBeq,
LCeq and LZeq are saved as a CSD file.
☆
Note:
After selecting the appropriate option, press <Enter>
on
the main screen
to manually save the data to the CSD file when the measurement is stopped.
CSD Log.Step
(CSD log step)
: