3.
Select Calibration from
the menu and press <Enter>
to
enter the By
Measurement
menu.
4.
Set the value for
Cal.Level
in the menu, for example to 93.8 dB. Calibration is
performed 5 seconds after pressing the <Start> button
.
5.
When the calibration is finished, the sound pressure meter updates the
calibration factor. Press the <Enter> button
to
accept the results.
6.
Go back to the
main screen
and press <Start/Stopp>
to
start the measurement.
The current measurement result in this example is 93.8 dB if the calibrator is in good
working order.
4.3 Measurement
There are 13 menu options in the
Measurement
menu. Use the arrow keys <▲>, <▼> to select
options and press <Enter> to move to the next
menu.