51
K000000000EJ 032433 210519
9.7
Configuration→Diagnostic
→
PARAMETER
DESCRIPTION
Displays information on the state and calibration of the electrochemical sensors:
Ok
No problem detected
absent
The sensor was not detected
err data
Memory data error of the sensor
unknown
It is necessary to update the FW of the device
err pos
The sensor has been installed in the wrong position
err cal
Calibration error (sensor not calibrated)
err curr
Currents outside the range
err cfg
Do not use this sensor as it has not been accepted on the screen "types of sensors".
Also, from this screen the user can access the identification data of the sensor: type, serial
number, date of manufacture and calibration. There are also the measured currents; in this way
it is possible to perform a quick diagnosis in the event of a malfunction.
Tests the tightness of the gas probe pneumatic path.
When instrument powers on, the firmware performs a full check on the physical Eff. tot of all
types of HW memories installed on the instrument, as well as on the integrity of the data stored
into them. Any issue is evidenced in the screen 'Memories Diagnostics'. Should this happen it is
advisable to turn the instrument off and then on again. In case the problem is permanent or
frequently recurring, the user should contact the Service Center reporting the error code shown
by the instrument.
In this submenu the user can temporarily turn the gas suction pump on or off. Also, it is possible
to view the actual flow rate of the pump in liters per minute. It will not be possible to turn off the
pump during an autozero cycle.
It is possible to make a recalibration of the instrument’s gas sensors with suitable known
concentration gas cylinders.
The access to the sensor recalibration is password protected, the password is ' 1111 '.
KEY
FUNCTION
Activate the context keys shown on the
display.
Returns to the previous screen.
CONTEXT KEY
FUNCTION
Selects the available parameters.
Enters in the selected parameter setting.
Selects the available parameters.
OK
◄
►
07/08/14
10:00
Sensors
Hardware
Gas probe
Pump
Diagnostic
On site cal.
Sensors
Gas probe
Hardware
Pump
On site cal.
Summary of Contents for E INSTRUMENTS 1500
Page 2: ......
Page 67: ...67 K000000000EJ 032433 210519 10 1 1 Memory Organization...
Page 166: ......
Page 168: ......
Page 169: ......