150
K000000000EJ 032433 210519
Turn on the instrument to check the new sensor works correctly through the menu "Sensor Troubleshooting".
It is normal if a newly installed sensor gives a 'current error': it is necessary to wait some time, so that the sensor
polarization can settle. The table here below shows the minimum settling time for each sensor.
Note:
(1) 2 hours settling time is recommended.
(2) 48 hours settling time is recommended; should the sensor be equipped with an external polarization battery, the
settling time is reduced down to 2 hours.
(3) 1/2-Hour settling time is recommended.
9
Reconnect the sensor (See point 3).
10
Close the back door of the sensor compartment again, and tighten screws again (See point 1).
8
Rotate the sensor clockwise until hearing a click (See point 4).
While rotating the sensor, take care not to exert any pressure onto the printed circuit
above: exert pressure onto the plastic body only.
CODE
DETECTED GAS
POSITION
SETTLING TIME
Flex-Sensor O
2
LL
Cod. AACSE43
O
2
Oxygen
S1
24 hours
(1)
Flex-Sensor O
2
Cod. AACSE15
O
2
Oxygen
S1
2 hours
(1)
Flex-Sensor CO+H
2
Cod. AACSE12
CO
Carbon Monoxide
S2
2 hours
(1)
Flex-Sensor CO high imH
2
Cod. AACSE20
CO
Carbon Monoxide
S2/S3/S4
2 hours
(1)
Flex-Sensor NO
Cod. AACSE10
NO
Nitrogen Oxide
S3
48 hours
(2)
Flex-Sensor NO
2
Cod. AACSE14
NO
2
Nitrogen Dioxide
S2/S3/S4
2 hours
(1)
Flex-Sensor SO
2
Cod. AACSE13
SO
2
Sulphur Dioxide
S2/S3/S4
2 hours
(1)
Flex-Sensor CO
100.000 ppm
Cod. AACSE17
CO
Carbon Monoxide
S2/S3/S4
2 hours
(1)
Flex-Sensor CO
20.000 ppm
Cod. AACSE18
CO
Carbon Monoxide
S2/S3/S4
2 hours
(1)
FLEX-Sensor CxHy
0-5.00% vol. referred to CH4
Cod. AACSE39
CxHy
unburnt hydrocarbons
S3/S4
1/2 hour
(3)
Flex-Sensor for gas leaks
Cod. AACSE19
Leak detector
Methane / LPG
S4
-
Flex-Sensor CO+H
2
low range
Cod. AACSE24
CO
Carbon Monoxide
S2
2 hours
(1)
Flex-Sensor NO low range
Cod. AACSE25
NO
Nitrogen Oxide
S3
48 hours
(2)
Flex-Sensor NO
2
low range
Cod. AACSE26
NO
2
Nitrogen Dioxide
S2/S3/S4
2 hours
(1)
Flex-Sensor SO
2
low range
Cod. AACSE28
SO
2
Sulphur Dioxide
S2/S3/S4
2 hours
(1)
Flex-Sensor CO
2
0 .. 20% v/v
Cod. AACSE21
CO
2
Carbon Dioxide
S3/S4
2 hours
(1)
Flex-Sensor CO
2
0 .. 50% v/v
Cod. AACSE47
CO
2
Carbon Dioxide
S3/S4
2 hours
(1)