BPLV700AV

2006-12-19

SATRON VG 

pressure transmitter

Figure 1-6

Figure 1-3

  Coupling mounting hole

Figure 1-7

1 Shut-off plug, AISI316 (Order No. M550405)
2 Shut-off screw G1
3  G1 welded process coupling

Figure 1-5

  Coupling shut-off plug

Figure 1-4

  Welding the coupling

3

10044004

DO  NOT  TOUCH!

10044006

10044108

Welding Sequence

45

.5

-0

.2

+

0

.5

1050354043

10044110

HEX 36

79

3

1

2

1.1.1 Installing welded process couplings

Mounting hole for coupling

• Make a 

Ø 

45.5 mm (+0.5/-0.2 mm) hole in the tank wall

or pipe, as shown in Figure 1-3.

Welding the coupling

These instructions apply to all welded couplings; welding
the G1  standard coupling is described here as an
example.
•  Place the coupling in the mounting hole as shown in Fig.
1-4. Then weld with several runs so as to prevent the
coupling’s oval distortion and tightness problems.
• The transmitter must be out of the coupling while the
coupling is welded.You can use the shut-off plug shown in
Fig. 1-5 to shut the coupling. The plug protects the
coupling’s sealing face and permits the starting of the
process without the transmitter.
• It is always recommendable to use the welding assistant
(M1050450) while welding the coupling to prevent any
distortions due to heat.
• Do not make weld grounding via any transmitter's body!

1.1.2 Mounting the transmitter on the coupling

Procedure

• Make sure that the coupling’s sealing face is clean.
• Remove the orange protective plug from the transmitter’s
diaphragm.

         Do not touch the diaphragm! Figure 1-6.

• Insert the transmitter  

in a straight line

  into the

coupling (Fig. 1-7), so that the guide groove on the
transmitter aligns with the stop pin on the coupling. The
transmitter settles into position when the groove and pin
are aligned, and will be prevented from rotating in the
coupling.

When inserting the transmitter, be careful not to
damage the edge of the sensor diaphragm on the
edges of the coupling or on the end of the stop pin!

• Lock the transmitter in position by screwing the hex nut
fully home. Finger tightness is sufficient to tighten the
sealing faces. However, we recommend final tightening
with a tool to eliminate the effect of vibration and other
such factors. Apply 60±20 Nm torque.

   Do not use sealing tape etc. on threaded
   connection!

1

3

5

2

4

6

Welding assistant M1050420

Summary of Contents for VG

Page 1: ...libration examples 4 CONSTRUCTION AND OPERATION 5 PARTS LIST DOCUMENTS TechnicalSpecifications BPLV700 InstallationandSetting UpInstructions BPLV700AV BPLV700AV Man rev 2 2006 12 19 Satron Instruments Inc P O Box 22 FIN 33901 Tampere Finland Tel int 358 207 464 800 Telefax 358 207 464 801 www satron com info satron com We reserve the right for technical modifications without prior notice HART is a...

Page 2: ... by installing heat insulation up to the sensor diaphragm SATRON VG pressure transmitter NOTE Dimensions are in millimeters 1300354152 installation with process coupling installation with PASVE mounting service valve Figure 1 1 Recommended mounting positions S Z Remote electronics connecting cable with protection hose codes L ja K Process temperature code H 40 Option L Option K Hex 36 160 170 Proc...

Page 3: ...e starting of the process without the transmitter It is always recommendable to use the welding assistant M1050450 while welding the coupling to prevent any distortions due to heat Do not make weld grounding via any transmitter s body 1 1 2 Mounting the transmitter on the coupling Procedure Make sure that the coupling s sealing face is clean Remove the orange protective plug from the transmitter s...

Page 4: ......

Page 5: ...ting service valve Thethree positionPASVE mounting servicevalvemakes the installation of the transmitter easy All necessary cleaning zero adjustment and replacement operations will be easy and fast to perform with PASVE without stopping the process Flange ANSI 3 150 lb ANSI 3 300 lb DN80 PN40 DimensionD 191 210 200 AISI316L diaphragm Viton O ring selection code 1 AISI316L diaphragm PTFE O ring sel...

Page 6: ...y converters The shield of the cable is grounded at the power supply end or according to the recommendations of the manufacturer of the used control system 1 Remove screw 2 Open Figure1 15 Removing the PLUG junction box 1 2 Open 0 5 1 5 mm 2 Pg9 10044093 10044092 BPLV700AV 2006 12 19 SATRON VG pressure transmitter 1300354153 10 12 15 20 25 30 35 0 250 500 750 1000 1400 Supply voltage V Load Ω 40 1...

Page 7: ...re 1 19 wiring housing M Figure 1 20 wiring housing N 1300354155 BPLV700AV 2006 12 19 SATRON VG pressure transmitter 1300354155 1300354155 Test Power Load mA Test mA Power Load Hart Test 3 2 1 1 3 60 5 DN UP PZ D RUN Z S 1 3 2 1 2 Power Load mA ...

Page 8: ...275 user interface After installing and connecting the transmitter connect the user interface to the transmitter The following menu is displayed 1 Measurement 2 Configuration 3 Information 4 Diagnostics To change the measuring range unit damping time constant to output mode linear square root select Configuration The following menu is then displayed 1 Range values 2 Detailed config To change the m...

Page 9: ... change the measurement unit select Range values The following menu is then displayed 1 LRV 2 URV 3 LSL 4 USL 5 Min span 6 Apply values To change the damping time constant select Output from theConfiguration menu The following menu is then displayed 1 Damping 2 Alarm current To change the output mode select Transfer function from the Configuration menu The following menu is then displayed 1 Lin fu...

Page 10: ...dows NT DMM Digital multimeter basic DCV accuracy better than 0 01 of reading for example Fluke 8840A Keithley 2000 Cal pressure generation and measurement device accuracy better than 0 03 of reading HART modem for example MicroLink Smart Modem Power supply Output 24 VDC Rout 10 ohm CAL BOX connection unit including precision current shunt 4 wire connections 100 ohm 0 01 overload protection circui...

Page 11: ... dev rev Software rev Hardware rev Descriptor Final asmbly num P SMART special Selection code P SMART Process seal Housing type Body Flange mater Diaph diaph mat diaphragm mater Diaphragm coating Extension material Fill fluid Explosion proof Diagnostics ALARMCPUEEER Device status flags ALARMADSENER ALARM ADTS ER Zero device status ALARM ADTH ER ALARMELDRER SCIERR ADERR HAFRERR HA DA ERR ADCEEERR L...

Page 12: ... Vol temp coeff Measuringpoint Deviceinfo Constructioninfo HART Tag Descriptor Message Date Manufacturer Model DevType Serialno Sensor no Device no Softwarerev Hardwarerev Selection code Process connection Housingtype Body flangemater Diaphragmmater Diaphragmmater Diaphragmcoating Extension material Ext lenght code Fill fluid Explosionproof Tempcode Process connect Universalrev Fld dev rev Polladd...

Page 13: ... the zero 13 3 CALIBRATION 3 1Adjustability Maximum span is 25 times the minimum span for SATRON VG transmitter Span adjustments is made from outside the housing under theprotectiverubbershield figure3 1 BPLV700AV 2006 12 19 SATRON VG pressure transmitter Use of selector switch RUN working position PZ Process value zero D damping adjustment S Span adjustment Z Zero adjustment DN Down UP Up Housing...

Page 14: ...urn the selector switch from RUN to position D 2 Turn the regulating switch about 20 so damping adjustment is activated Turn the regulating switch to desired value of damping 0 s on the left side 60 s in the right side 3 Turn the selector switch from position D to position RUN Figure 3 4 Damping adjustment Figure 3 5 Process value zero PZ 3 3 Calibration examples The calibration stand shown in Fig...

Page 15: ...ut is closest to 20 mA adjustment range on fine adjustment range is 0 75 of span and speed of adjustment is 2 5 of span s 4 Turn the selector switch from position S to position RUN Apply zero pressure Repeat the adjustments to achieve the desired accuracy Sensor Module Analog to Digital Signal Conversion Sensor Temperature Non Volatile Memory Correction Coefficients Module Information Sensor Micro...

Page 16: ... bridge ElectronicsModule The electronics module converts the process pressure signal from the sensor module to 4 20 mA output signal The conversion can be made in linear square root or inverted mode or it can be done through user selectable pressure output point pairs 2 16 points Transmitters provided with own display code N is equipped with operating keys that allow you to define the transmitter...

Page 17: ...Figure 5 2 Parts list Enclosure with remote electronics 17 1300354163 BPLV700AV 2006 12 19 SATRON VG pressure transmitter 7 14 15 16 17 18 13 10 9 8 4 3 2 1 11 ...

Page 18: ...e 5 3 Parts list Enclosure M housing with terminal board Figure 5 4 Parts list Enclosure N housing with display 1300354161 BPLV700AV 2006 12 19 SATRON VG pressure transmitter 1300354162 20 19 1 13 21 22 1 23 13 ...

Page 19: ... M3 x 4 VSM 13302 Zne Protection cup housing H M and T Protection cup housing N Coupling Mounting clamp Support plate Hex nut M8 SFS2067 A4 Mounting bracket S O ring 42x2 FPM Viton Cover M Seal N Silicone rubber Cover N Cylinder head screw S M4x8 SFS2176 A4 Name 1 2 3 4 7 8 9 10 11 Number 19 Notes BPLV700AV 2006 12 19 SATRON VG pressure transmitter Order number Number 13 13 14 15 16 17 18 19 20 21...

Page 20: ... FIN 33901Tampere Finland Tel int 358 207 464 800 Telefax 358 207 464 801 BPLV700AV 2006 12 19 SATRON VG pressure transmitter MEETS THE COUNCIL OF THE EUROPEAN UNION DIRECTIVE 89 336 EEC FOR ELECTROMAGNETIC COMPATIBILITY REQUIREMENTS ...

Reviews: