Figure 4-1 Functional construction of the SATRON VG pressure transmitter
15
BPLV700AV
2006-12-19
SATRON VG
pressure transmitter
Figure 3-6 Set the lower range-value
Figure 3-7 Set the upper range-value
Measuring range: 0...300 kPa (range 5 transmitter).
Span: 300 kPa
Procedure
-
Apply zero pressure.
1. Turn the selector switch from position RUN to
position Z.
2. Turn the regulating switch about ±20° so adjustment is
activated.
3. Turn the regulating switch to a point where output is
closest to 4 mA. (adjustment range on fine adjustment
range is ±0.75% of span and speed of adjustment is
±2.5% of span / s)
4. Turn the selector switch from position Z to position
RUN.
-
Apply full-span pressure.
1. Turn the selector switch from position RUN to
position S.
2. Turn the regulating switch about ±20° so adjustment is
activated.
3. Turn the regulating switch to a point where output is
closest to 20 mA. (adjustment range on fine adjustment
range is ±0.75% of span and speed of adjustment is
±2.5% of span / s)
4. Turn the selector switch from position S to position
RUN.
-
Apply zero pressure.
-
Repeat the adjustments to achieve the desired
accuracy.
Sensor Module
Analog-to-Digital
Signal Conversion
Sensor
Temperature
Non-Volatile Memory
- Correction Coefficients
- Module Information
Sensor Microcomputer
and Memory
- Sensor Linearization
Electronics Module
Non-Volatile Memory
- Range Values
- Transmitter Configuration
Output Microcomputer and Memory
- Rerange
- Damping
- Diagnostics
- Engineering Units
- Communication
Keyboard
LCD-Display
Digital-to-Analog
Signal Conversion
Interface
Stage
- Failure
protection
- Test
Connections
- Cable
Connections
Analog Output
Signal to
Control
System
Pressure
Piezoresistivity
Sensor
- User function
HART® Digital Comm.
Module
Temperature
1300354154
1300354154
1300354156
±20°
±20°
Ha
rt
®
Test
3
2
1
1
3
60
.5
DN UP
PZ
D
RUN
Z
S
1
3
60
.5
DN UP
PZ
D
RUN
Z
S
Ha
rt
®
Test
3
2
1
1
3
60
.5
DN UP
PZ
D
RUN
Z
S
1
3
60
.5
DN UP
PZ
D
RUN
Z
S
Summary of Contents for VG
Page 4: ......