Minimum Load
The minimum load required for
initialization of the weighing platform
is configured in the operating menu
under:
Appl
Z
nM
:
3.6.
Once the limit is exceeded by the load,
initialization can begin. If the load is
too light, the following will occur when
you try to save a value:
– The error code
Inf 29
is displayed
– The weighing platform is not initialized
– The preset calculation factor is saved
The minimum load required for auto-
matic taring of the container weight on
the platform (“autotare first weight") is
configured in the operating menu
under:
Appl
Z
nM
:
3.5.
You can choose from 10 settings,
ranging from
1 digit
to
1000 digits
Example: If the scale interval (d) is
1000 g and the minimum load is set
to 1000 digits (=1000 scale intervals),
a load of at least 1000 g is required
for autotaring.
Resolution for Calculation of
Reference Value
The resolution applied for calculating
the reference value is defined in the
operating menu under:
Appl
Z
nM
:
3.9.
The resolution for calculating the
reference sample is increased if
“+1 decimal place", “+2 decimal places"
or “With internal resolution" is selected.
With the “+1 decimal place" setting,
the net value is determined to one
additional decimal place (i.e., display
accuracy
+
10); “+2 decimal places"
increases display accuracy
+
100, and
so on up to the maximum resolution
available.
Parameter for Saving Weight Values
The reference weight is saved when
the platform has stabilized. “Stability"
is defined as the point at which
fluctuation of a measured value lies
within a defined tolerance range.
The narrower the tolerance range, the
more stable the platform is at stability.
In the operating menu, under:
Appl
Z
nM
:
3.11.
you can define whether the value
is saved when “standard stability" is
reached, or only at “increased stability"
(narrower tolerance range). If you select
“At increased stability," the reference
weight saved will be more accurate and
the results more reproducible, but the
response time of the weighing instru-
ment might be longer.
Decimal Places for Display of Results
In neutral measurement, not only
whole numbers but also decimal
numbers (for example, 1.25
o
electrical
cabling) can be displayed.
The number of decimal places displayed
in neutral measurement is configured
in the operating menu under:
Appl
Z
nM
:
3.10.
62
Summary of Contents for Midrics 1
Page 1: ...Operating Instructions Sartorius Midrics 1 Midrics 2 Models MIS1 MIS2 Indicators 98648 014 89 ...
Page 92: ...92 Dimensions Scale Drawings All dimensions given in millimeters 92 ...
Page 97: ...97 C ...
Page 98: ...98 ...
Page 99: ...99 ...
Page 100: ...100 ...
Page 101: ...101 ...
Page 102: ...102 ...
Page 104: ...104 ...