![Sartorius MA45 Operating Instructions Manual Download Page 30](http://html1.mh-extra.com/html/sartorius/ma45/ma45_operating-instructions-manual_1199628030.webp)
Step
Key (or Instruction)
Display/Printout
--------------------
13.08.2001
16:01
Model
MA45-000230V
Ser. no.
13206969
Ver. no.
01-43-01
ID
--------------------
Prg. 1
Heating
STANDARD
Fin. temp.
130 ’C
Stdbytemp.
OFF
Start
W/STABIL.
End
AUTOMATIC
IniWt +
2.036 g
The current moisture loss is displayed
after the data is printed.
The drying program shuts off
automatically once no further
weight loss is detected (in this
example, after 5.2 minutes)
Then the footer of the analysis
--------------------
record is printed
FinWt +
1.814 g
5.2+
18.90 %L
Name:
--------------------
30
0
05
PROGRAMM
P1 130
o
C Auto.
37
o
C 0.8min
S
%L
CANCEL
SETUP
1
8
90
PROGRAMM
P1 130
o
C Auto.
128
o
C 5.2min
S
%L
NEXT
SETUP