![Sartorius MA45 Operating Instructions Manual Download Page 14](http://html1.mh-extra.com/html/sartorius/ma45/ma45_operating-instructions-manual_1199628014.webp)
Setting the Device Parameters
(
DEVICE
)
Purpose
Device configuration, i.e., to meet
individual requirements by selecting
predefined menu parameters in the
Setup menu. You can block access to
the menu by assigning a password.
Features
The device parameters are combined in
the following groups:
– Password to the Setup menu
– Interface
– Display contrast
– Acoustic signal
– Factory settings
You can display, enter or change the
following parameters:
Password
– Password for access to the device
parameters and isoTEST functions
in the Setup menu, and to drying
programs in the Program menu
Interface
– SBI operating mode
Simple record of analysis results for PC
or external printer, factory setting for
the YDP03-0CE printer
Format: baud rate, number of data bits,
parity, stop bits, handshake
– xBPI operating mode
Function-oriented interface with trans-
parent data transmission
Network address: enter a number from
0 to 31; factory setting: 0
Display contrast
– Contrast/angle of the display (enter a
number from 0 to 4; factory setting: 2)
Acoustic signal
– Acoustic signal on or off
Factory Settings
Parameters: The factory-set configura-
tions are identified by an “
ο
” in the list
starting on page 17.
Preparation
Display existing device parameters
§
Select the Setup menu:
Use the
x
or
y
key to select
SETUP
in the Function line and press
u
to confirm
> SETUP is displayed:
§
Select Device parameters: Press the
y
key and press
u
to confirm
If no password has been assigned,
anyone can access “SETUP: Device
parameters” without entering a
password
If a password has been assigned:
> The password prompt is displayed
$
If the Setup menu is protected by a
password: Use the
x
and
y
keys
to enter the password (see page 9 for
details) and press
u
to confirm
§
Confirm the password and display
“Device parameters”:
Press the
u
key
> Device parameters are displayed:
Entering or Changing the Password
– The password (up to 8 characters)
protects the following functions:
– Setup: Device parameters
– Program: Drying programs
– Setup: isoTEST functions
§
Display Device parameters (see
“Preparation > Display existing device
parameters” above)
§
Write down the password here for easy
reference:
Password = . . . . . . . . . . . . . . . . . . . . . .
If you assign a password and then
forget what it is:
$
Enter the General Password
(see Appendix)
$
Confirm the password and display
device parameters:
Press the
u
key
> Parameters are displayed (see above)
§
Configuring device parameters:
Select “Password” (if configured) and
confirm: Press the
u
key
>
Password:
and any existing password
are displayed:
§
New password: Use the
x
and
y
keys to enter the new password (up to
8 characters) and press
u
to confirm
If no entry is shown under “Password”,
then no password has been stored
§
To confirm your input:
Press the
u
key
§
Exit the Setup menu:
Press the
c
key twice
14