![Sartorius MA45 Operating Instructions Manual Download Page 15](http://html1.mh-extra.com/html/sartorius/ma45/ma45_operating-instructions-manual_1199628015.webp)
Additional Functions
Print out parameter settings:
§
If “Device parameters” are displayed:
Press the
r
key
> Printout (example)
Lines with more than 20 characters
are truncated
--------------------
SETUP
DEVICE
--------------------
Interface
SBI
Baudrate
1200 baud
Number of data bi
7 data bits
Parity
Odd
Number of stop bi
1 stop bit
Handshake mode
Hardware handshake a
fter 1 char
Displaycontrast
2
Acoustic signal
On
--------------------
15
Reset the device parameters to the
factory settings:
§
Select the Setup menu:
Use the
x
or
y
key to select
SETUP
in the Function line and press
u
to confirm
> SETUP is displayed:
§
Select Device parameters:
Press the
y
key and press
u
to confirm
$
If a password has been assigned: enter
the password as described above
§
Confirm the password and display
“Device parameters”:
Press the
u
key
> Device parameters are displayed
§
Select
Factory settings
:
Press the
y
key 4 times
§
Confirm
Factory settings
:
Press the
u
key
$
To cancel the “Reset” operation at
this point: Press the
c
key
§
Select
Yes
and confirm.
Press the
y
key and then
press the
u
key
§
Exit the Setup menu:
Press the
c
key three times