![Sartorius MA35 Operating Instructions Manual Download Page 38](http://html1.mh-extra.com/html/sartorius/ma35/ma35_operating-instructions-manual_1199626038.webp)
38
Overview
Specifications
Weighing capacity (Max)
35 g
Typical reproducibility of the
weighing system
Starting at an initial sample weight of approx . 1 g: ± 0 .2%
Starting at an initial sample weight of approx . 5 g: ± 0 .05%
Readability
1 mg; 0 .01%
Typical sample quantity
5 – 15 g
Display of results
% moisture
% dry weight
% ratio
g residual weight
Shutoff parameter
optional:
– Fully automatic
– Manual shutoff
– Timer settings 0 .1–99 min
Sample heating
Even sample heating capability with an IR dark radiator
(metal tube heater)
Access to sample chamber
– through hinged, flip-up cover with wide opening angle
– with viewing window
Draft shield
Integrated draft shield
For conformity with FDA/HACCP regulations Aluminum panels (in place of glass panels)
Operating temperature range and setting
40°C to 160°C (104°F to 320°F), adjustable in 1°C increments
Operator guidance
Symbols representing corresponding functions
Heating program
Standard drying
Measuring program
1 program saved in a non-volatile memory (freely selectable
parameters)
Memory for data storage
Results are saved until the start of the next measurement
Log printout
– Inalterable standard configuration GLP report for
measurement results/calibration/adjustment
– Short report for measurement results minimizing
paper usage
– Using an optional external printer
Data interface
RS232 C-S/VA24-V28, 7 data bits (ASCII) 1 parity bit
Monitoring and control of the accuracy of
the analyzer as inspection, measuring and
test equipment
External calibration with optional calibration weights;
Temperature adjustment with an optional adjustment set
Housing dimensions (W × D × H)
224 × 366 × 191 mm
Net weight
approx . 5 .8 kg
Power requirements (supply voltage)
230 V or 115 V (depending on the model) (–15% … +10%)
Frequency
50 – 60 Hz
Fuses
2 (zero conductor/phase), 6 .3 A, time-lag (slow-blow),
5
+
20 mm (internal)
Power consumption
max . 400 VA