31
Step
Key (or instruction) Display
The prompt for calibration
weight is displayed
8 . Open the hinged cover
9 . Place the 30-weight on
the weighing system
Minus sign –: weight value too low
Plus sign +: weight value too high
No sign: weight value OK
The weight unit symbol (g) is
displayed at the end of adjustment
Printout after calibration and
--------------------
adjustment
23.03.2017 10:51
*)
*) Date and time included only
Model MA35M-230N
if a Sartorius printer model
Ser. no. 99992581
YDP20-0CE is used
Ver. no. 00-33-01
ID
--------------------
External calibration
W-ID
Nom. + 30.000 g
Diff. + 0.001 g
External adjustment
Diff. + 0.000 g
completed
--------------------
Name:
--------------------
10 . Unload the analyzer
Close the cover
11 . Quit calibration/adjustment
c
Summary of Contents for MA35
Page 2: ......