Example: setting the following drying parameters
Heating temperature:
130°C
Start of analysis:
without stability after the cover is closed
End of analysis:
after 10 minutes
Display mode for result: moisture
Step
Key (or instruction) Display
1 . Turn on the analyzer
e
Self-test runs
2 . Select PRG: drying program
parameters
3 . Confirm PRG
105
°
C
(the previously set temperature
is displayed; in this example, 105°C)
4 . Set the heating temperature
repeatedly
130
°
C
(in this example: 130°C)
5 . Confirm heating temperature
0.0
min
(the previously set analysis time
is displayed; in this example, 0 .0 min)
6 . Set the parameter for the
repeatedly
10
min
end of analysis; in this example,
10 minutes)
7 . Confirm the “end of analysis”
parameter
8 . Select the result display mode
or
%M
(in this example, moisture)