![Sartorius MA35 Operating Instructions Manual Download Page 24](http://html1.mh-extra.com/html/sartorius/ma35/ma35_operating-instructions-manual_1199626024.webp)
24
Drying Parameter Features
– Heating temperature set to defined value
during measurement
Start of analysis
– With stability after the
key is
pressed:
When START is shown in the function
line and you press
to confirm, the
initial weight is stored at stability
regardless of whether the cover is open
or closed .
Measurement begins as soon as the cover
is closed .
– Without stability after the cover is
closed:
A symbol shown in the graphic symbol
display prompts you to close the cover
once the initial weight condition is met .
The initial weight is stored without
stability as soon as the sample chamber
is closed, and analysis begins .
End of Analysis with Shutoff
Parameters
– Fully automatic mode
– Timer mode
Fully automatic mode:
Use the fully automatic mode when
loss of weight on drying follows a
clearly delineated curve which can be
unambiguously evaluated (see below) .
Weight
Moisture (evaporation)
Automatic shutoff
Time
Timer mode:
The analysis ends as soon as the
specified time has elapsed .
Display Mode for Result
The following units can be selected for
displaying analysis results:
– Moisture
%M
– Dry weight
%S
– Ratio
%MS
– Residual weight g
Print Intermediate Results
Intermediate results can be printed
either at user-definable intervals or by
pressing the
r
key .
Summary of Contents for MA35