30
External Calibration and Adjustment with a Factory-Defined Weight
Externally calibrate and adjust the weighing system using a 30-g calibration weight
Step
Key (or instruction) Display
1 . Turn on the analyzer
e
Self-test runs
2 . Select CAL for calibration/
adjustment
3 . Confirm CAL
4 . Confirm again when
Pb
is shown
5 . Tare the weighing system
6 . Select CAL again
7 . Confirm CAL