Error messages are shown in the text line or measured value line of the display. 

Display 

Cause 

Solution

No segments shown 

No AC power is available 

Check the AC power supply  

 

AC adapter not plugged in 

Connect AC adapter to the  

 

 

wall outlet (mains) 

 

Fuses are defective 

Replace the fuses 

 

The analyzer is switched off  

 Press the power switch (on/off)  
on the back panel

L / ERR 54

 

Load is below the weighing range 

Place a load on the weighing module 

 
 

Sample retainer not in place 

 Place the sample retainer on the  
weighing module

H

 

The load exceeds the weighing  

Unload the weighing module 

 

module capacity

ERR 02 

Calibration/adjustment condition 

Calibrate only when zero is displayed 

 

not met, e.g., 

 

– not tared  

Tare the weighing module 

 

– the sample retainer is loaded 

Unload the sample retainer

ERR 03

 

Calibration/adjustment could not be  

Allow the scale to warm up  

 

completed within the required time period 

again and repeat the adjustment

expected xxx 

Wrong format 

Enter data correctly 

 

Wrong format

bad xxx/illegal xxx 

Outside the range 

Enter data correctly 

Not permitted

must not be more than xxx 

Outside the range 

Enter data correctly 

Too high

must not be less than xxx 

Outside the range 

Enter data correctly 

Too low

Value exceeds display 

Value has more characters 

Change configuration.  

 

than the display can show 

See the chapter entitled “Configuration,”  

 

 

 under “Configuring Parameters in the  
Setup Menu”

ERR Config. data

 

Checksum errors; may indicate change  

Switch the moisture analyzer off  

 

in versions (e.g., after software update) or 

and on again  

ERR Adjust. data

 

dead battery 

 

ERR Printout config. 
 
ERR DB user 
 
ERR DB programs 
 
ERR DB results

Messages

Messages are shown on the display at the time the event triggering the message occurs (covers other displayed data). For details on 
 messages, see the chapter entitled “Operating Design.”

Troubleshooting Guide

Summary of Contents for LMA200PM

Page 1: ...Operating Instructions Sartorius Microwave Moisture Analyzer Model LMA200PM 98648 015 60 98648 015 60...

Page 2: ...disinfect Symbols The following symbols are used in these instructions Instructions for proper operation Instructions for exceptional cases The outcome of an operating step Warning hazard Intended Use...

Page 3: ...re the cooling fans are always operational In the event of defect switch off the moisture analyzer and inform service personnel Warnings and Safety Precautions If you use cables purchased from another...

Page 4: ...mber Do not use a sample container that prevents the cover from closing completely The moisture analyzer must not be operated if the sample cham ber is not completely closed and sealed by the analyzer...

Page 5: ...secondary raw material Equipment Supplied LMA200PM microwave moisture analyzer model LMA200PM 000EU or LMA200PM 000US Adapter with sample retainer Power cord Glass fiber pads 80 Printer paper 1 roll P...

Page 6: ...e the protective grounding effect is not neutralized by use of an extension cord that lacks a protective grounding conductor and a fuse with maximum 16A The power plug or other suitable device for dis...

Page 7: ...ve function Navigating the menu Returns to previous menu level During input Cancels input Confirm During measurement Activates the selected function Navigating the menu Shows next menu level During in...

Page 8: ...Measured value result line This line shows the weight readout with the weight unit at stability or a calculated value a plus or minus sign and the weight unit displayed Text lines The text lines show...

Page 9: ...Tap the second touchscreen button from the left the arrow pointing towards the left 9 Entering data using the touchscreen keypad If no is shown the value can be entered or edited using the touchscree...

Page 10: ...w of the Moisture Analyzer on page 4 Built in printer Ethernet interface RS 232 interface Built in Printer You can configure printer settings and printout formats including settings for ISO GLP confor...

Page 11: ...ap the Login button First login Admin Administrator Subsequent login list of all users Tap the First login Administrator line Admin PIN Subsequent login Name PIN Tap the name PIN line Enter your PIN F...

Page 12: ...this example Jones Group Operator and press the key PIN 4 Select the Supervisor group Tap the button Group Supervisor and press the key 5 Open the PIN input window Tap the button and press the key 6 E...

Page 13: ...Results are displayed as follows Each dataset is shown on two lines First line Internal number program name date time Second line User name sample ID result with unit and analysis time There are a num...

Page 14: ...ic printing Headers 1 to 3 Print formats You can set the following parameters for printing data records and statistics Data record header Data record intermediate results Data record footer Statistics...

Page 15: ...erature Info Shows information about the LMA200PM the weighing module and the databases To print an informational screen Press the Print key 15 Overview of User Privileges Group Administrator Supervis...

Page 16: ...page 17 Results Sample ID See page 19 Evaluation of results Printing parameters Automatic printout Header 1 Header 2 Header 3 Print formats Header Intermediate results Footer Statistics See page 19 C...

Page 17: ...nfo only Device name Alphanum Use DHCP Option IP address nnn nnn nnn nnn Subnet mask nnn nnn nnn nnn Standard gateway nnn nnn nnn nnn Remote access for VNC client nnn nnn nnn nnn Operating parameters...

Page 18: ...formats Setup Configuration Device Serial Built in On interfaces printer Off RS 232 Off Printout Protocol XON XOFF parameters RTS CTS 2 wire No protocol Baud rate 300 bd 600 bd 1 200 bd 2 400 bd 4 80...

Page 19: ...termediate Blank line Header 1 results Header 2 Date and time Header 3 Time GLP header ID1 Date and time ID Time Analysis time Result Operator ID1 Footer Blank line ID2 Program name Date time Drying p...

Page 20: ...0 1111 to 9999 repeat PIN input User privileges SET Calibration Supervisor SET Manage programs SET Eval of results SET Print param SET Printout formats SET Date and time SET Serial ports SET Ethernet...

Page 21: ...range Analysis mode Analysis mode Standard Addition of water Concentration mg l Drying process Standard Level 2 to 100 2 stages Level 1 2 to 100 30 to 100 Time 1 0 0 to 10 0 min Level 2 2 to 100 30 to...

Page 22: ...01 0 00 Result display mode Moisture m Conversion factor 1 0000 to 2 0000 Solids S Conversion offset 50 000 to Volatile V 50 000 ppm moist ppmM ppm solids ppmS Ratio MS Weight loss mg Residue g Create...

Page 23: ...nto stages with different microwave output in each stage Note The time of the last stage is the deter mined by the shutoff parameter in all drying processes Standard The microwave output remains const...

Page 24: ...oist ppmM ppm solids ppmS Ratio MS Weight loss mg Residue g You can define the following for the Moisture M and Dry weight S display modes Conversion factor Conversion offset If one of these parameter...

Page 25: ...th information is shown briefly i Loading The display s info line shows LMA200PM Rel 01 00 00 Prg Ketchup 5 Begin moisture analysis Tap the Analyze button 28 11 2007 14 41 38 0 000 g Prg Ketchup Mode...

Page 26: ...11 2007 16 41 46 Sartorius Microwave Moisture Analyzer LMA200PM 000EU Rel 01 00 00 Serial no 12345678 Prg Ketchup Mode Standard Process R Standard Level 30 End Automatic Initial weight 2 125 g During...

Page 27: ...on the weighing module Remove any pad or sample from the sample retainer 5 Tare the weighing module Tap the Tare button Status line yellow background Taring the weighing module 6 Calibrate the weighin...

Page 28: ...rrent program or Create new program the bar graph shows abso lute values 0 to 70 g Checkweighing graph If Minimum maximum or Setpt tolerance is selected for the Initial weight parameter the bar graph...

Page 29: ...ions Menu lines are not touch sensitive Navigation and selection are possible only using the touchscreen buttons at the bottom of the display Press the key to insert selection Press the key to return...

Page 30: ...orius cables with order number 7357312 or 7357315 RS 232 Interface Type of interface Serial interface Operating mode Full duplex Standard RS 232 Protocol XON XOFF RTS CTS 2 wire or none Transmission r...

Page 31: ...egardless of the labels in the above example the arrow keys shown on the touchscreen buttons at any given point The following equivalents also apply F5 Back F12 Print Enter Confirm Arrow keys Like the...

Page 32: ...IP address of computer 192 168 178 20 IP address of LMA200PM 192 168 178 21 The settings for subnet mask and standard gateway should be the same as those in the local computer These settings are confi...

Page 33: ...ot be Allow the scale to warm up completed within the required time period again and repeat the adjustment expected xxx Wrong format Enter data correctly Wrong format bad xxx illegal xxx Outside the r...

Page 34: ...fter analysis Open the cover Use the forceps to grasp the glass fiber pad and carefully remove it from the sample chamber Replacing the Printer Paper Press down briefly on the cover of the printer com...

Page 35: ...particular the sealing surfaces between the upper and lower parts of the sample chamber is not permitted as this will damage the PTFE coating If any part of the sealing surfaces between the upper and...

Page 36: ...30 V 50 Hz 1200 VA LMA200PM 000EU 115 V 60 Hz 1200 VA LMA200PM 000US Fuse 2 10 A T 5 20 mm Ambient conditions Indoor use only Altitude up to 2000 m Operating temperature range 5 C to 40 C 41 F to 104...

Page 37: ...37 All dimensions given in mm Dimensions Scale Drawings...

Page 38: ...38...

Page 39: ...orms are used to facilitate legibility in these instructions and always simultaneously denote the other gender as well Copyright notice This instruction manual including all of its components is prote...

Reviews: