![Sartorius LMA200PM Operating Instructions Manual Download Page 26](http://html1.mh-extra.com/html/sartorius/lma200pm/lma200pm_operating-instructions-manual_1199622026.webp)
Step
Key (or instruction)
Display/output
9. Open the hinged cover
Open the LMA200PM
Prompt (green background):
Remove the pads
Add sample to pad
10. Add sample
Trickle ketchup onto glass fiber pad
Spread ketchup evenly
Place the second pad on the ketchup
11. Insert sample
Place pads with ketchup on the
If the sample weight is within the
sample retainer
tolerance limits:
An acoustic signal (beep) is emitted
The bar graph is green
Close cover or press Start
If the sample weight is not within
tolerance limits:
No acoustic signal is emitted
The bar graph is orange or red
Change sample amount as needed
Remove some ketchup from the pad, or
add more ketchup
12. Continue analysis
Close the analyzer cover
The status line briefly shows
the following:
Determining initial weight value...
The header is printed. Example:
–––––––––––––––––––––––––––––––––––––––––
28.11.2007 16:41:46
Sartorius Microwave Moisture Analyzer
LMA200PM-000EU Rel. 01.00.00
Serial no. 12345678
–––––––––––––––––––––––––––––––––––––––––
Prg. Ketchup
Mode: Standard | Process: R Standard
Level: 30 %
End: Automatic
Initial weight: + 2.125 g
–––––––––––––––––––––––––––––––––––––––––
During analysis, the status line shows
the following (yellow background):
Analysis in progress...
At the end of analysis, the status line briefly
shows the following (yellow background):
Determining final weight...
The analysis result is printed. Example:
–––––––––––––––––––––––––––––––––––––––––
00:40 + 0.752 g + 64.63 %M
–––––––––––––––––––––––––––––––––––––––––
After the end of analysis, the prompt line
shows the following (green background):
Run completed; remove sample
The bar graph line shows the percentage
of solids in the initial weight.
The measured value line shows, for example:
Moisture
+ 64.63 %M
13. Remove the sample
Raise the analyzer cover
The cooling fan continues to run.
Remove the pads with the sample.
The cooling fan shuts off automatically.
14. Begin next analysis
Tap the
Analyze
button
Continue from Step 4 above.
Cooling fans switch off.
15. End the analysis series.
Press the
key
Starting screen