![Sartorius LMA200PM Operating Instructions Manual Download Page 15](http://html1.mh-extra.com/html/sartorius/lma200pm/lma200pm_operating-instructions-manual_1199622015.webp)
Restore factory settings:
You can restore factory settings sepa-
rately for any of the following areas:
– Program database
All drying programs are deleted. One
program, called “Default” (short name),
is created with the factory-default
parameters.
– Results database
All results are deleted.
– User database
All user accounts are deleted. One user,
called “Admin,” is created in the Admin-
istrator group with the PIN “9999.”
– Printing options:
All printing parameters and printout
options are reset to the factory defaults.
– Device configuration
The following parameters are reset:
– Format for display and printout of
date and time
– Built-in printer
– Function of the RS-232 interface
The following parameters are not reset:
– Clock
– RS-232 data interface parameters
– Ethernet network parameters
– Operating parameters (language,
PS/2 keyboard, keyclick duration)
– Display
Additional configurations:
You can configure the following:
– Info line for starting screen
The text you enter here (up to 50 char-
acters) is displayed directly below the
weight readout in the initial display.
If you do not enter any text here, none
is shown in the initial display.
– Maximum magnetron temperature
You can enter a temperature from 50 °C
to 80 °C. No analysis can be performed
if this temperature is exceeded. If you
tap the
Analys.
button while this
function is blocking analysis, the mes-
sage “Cooling down” is displayed until
the analyzer has returned to operating
temperature.
Info
Shows information about the
LMA200PM, the weighing module and
the databases.
§
To print an informational screen:
Press the
key.
15
Overview of User Privileges
Group
Administrator Supervisor
Operator
Operator
with PIN=0
SETUP Security management
(+)
(–)
(–)
(–)
User login/logout
(+)
(+)
(+)
(+)
SETUP Manage programs
+
+
–
–
Call programs
(+)
(+)
(+)
(+)
Perform moisture analysis
(+)
(+)
(+)
(+)
SETUP Evaluate results
(+)
(+)
(+)
(+)
SETUP Info
(+)
(+)
(+)
(+)
SETUP Calibration
+
–
–
(–)
SETUP Configuration
DEVICE Date and time
+
+
–
(–)
DEVICE Serial interfaces
+
+
–
(–)
DEVICE Ethernet
+
–
–
(–)
DEVICE Operating para
+
–
(–)
DEVICE Display settings
+
+
+
(–)
Results
+
+
–
(–)
Print Parameters
+
+
–
(–)
Print formats
+
+
–
(–)
Weighing module settings
+
+
–
(–)
+:
Permitted
(+):
Always permitted; cannot be modified
–:
Not permitted
(-):
Not permitted; cannot be modified
Notes:
– Users with operator privileges and PIN=0 can run the permitted functions without
entering a PIN. Login is solely for identification purposes in documentation of results.
– The user account with the name “Admin” in the Administrator group cannot be deleted.
We recommend changing the PIN for this user account so that it no longer uses the
factory-set PIN listed in this manual.
– Administrators can configure access controls for the Supervisor and Operator groups.
– If you forget the administrator PIN, tap the “Service” touchscreen button. This generates
a random number; contact the Sartorius service hotline to receive the temporary PIN
assigned to this number.
– Tap the Logout touchscreen button and enter your PIN to log out. A message opens
to confirm logout.
– Users are logged out automatically if the LMA200PM is disconnected from power.
Pressing the on/off key to set the analyzer in standby mode does not log the user off.