Scale Operating Program 
 
 

The scale operating program lets you adapt your scale to various am-
bient conditions and to different weighing requirements, and select 
weight units commonly used in your country. 
At the factory, we have set the codes for a Standard program which is 
protected by a locking function to prevent accidental changes. 
 
The "

menu code

" contains the information of the operating program, 

lt consists of three digits, known in "computerese" as the page  
(1st digit), the line (2nd digit) and the word (3rd digit). 
 

Reading a Program Menu Code (List Mode) 

 
How to access the menu of the scale operating program: 
 
With the display turned off (

STANDBY

 state), hold down the tare con-

trol 

(6)

 and briefly press the ON/OFF key 

(8)

. Upon completion of the 

automatic self-test, release the tare control as soon as "

CH5

" is dis-

played. The Status of the scale operating program is indicated in the 
weight display: 
 
"

L

" Stands for the list mode. In this mode, you can read a menu code 

setting, but you cannot change it. 
 
 
 
 

Changing a Program Menu Code (Change Mode) 

 
lf you wish to change a program menu code, you must first unlock the 
menu access switch to access the program menu. 
 
To do so, remove the protective cover located on the right of the dis-
play unit, and slide the menu access switch 

(5)

 in the direction of the 

arrow. 
 
The display will indicate "

C

", which Stands for the change mode, 

meaning you can now change the menu code settings. 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for IB 16000 S

Page 1: ...Sartorius Industry IB 16000 S IB 31000 P IB 31 Electronic Precision Scales Installation and Operating Instructions 98648 000 60...

Page 2: ...r for scale base 6 Tare control 2 Level indicator only for IB 16000S 7 CAL key 3 Power receptacle 8 ON OFF key 4 Leveling foot 9 Weight display 5 Menu access switch 10 Manufacturer s label for scale o...

Page 3: ...ltage Selection 16 Safety Precautions 17 Connecting Electronic Devices Peripherals 17 Leveling the Scale using the Level Indicator 17 Operation 18 Turning the Display On and Off 18 Self Test 19 Weighi...

Page 4: ...quipment delivered Electronic Precision Scale series IB is manufactured and tested in compliance with the following DIN VDE regulations DIN IEC 348 VDE 0411 Safety requirements for electronic measurin...

Page 5: ...s been designed to ensure that the scale will not be damaged even if it is dropped from a height of 80 cm about 32 inches After unpacking the scale please check it immediately for any visible damage a...

Page 6: ...ent supplied includes the components shown on the left Scale with platform and display unit AG adapter Dust cover 3 Allen wrenches Retainers for mounting the display 2 M 4 x 8 screws and washers for m...

Page 7: ...rea lf you need to transfer the scale to a warmer area make sure to condition it for about 2 hours at room temperature leaving it unplugged Afterwards ifyou keep the scale connected to AC power the co...

Page 8: ...as follows base mounted on the front of the scale factory mounted base mounted on the side of the scale see next page for direc tions mounted as a remote display unit see page 11 mounted on a support...

Page 9: ...rests on the platform cover Remove the screws from the base plate using the Allen wrench sup plied see arrows and then remove the base plate Unscrew the four screws of the angular cover plate and rem...

Page 10: ...y as shown in the diagram on the left Make sure that the rubber seal see arrow is not damaged Reposition the base plate and fasten it in place by tightening the three M3x8 screws see the large arrows...

Page 11: ...platform cover Remove the screws from the base plate see arrows using the Allen wrench supplied and then remove the base plate Unscrew the four screws of the angular cover plate and remove the angular...

Page 12: ...ead the cable through one of the three openings in the scale hous ing Make sure that the rubber seal see small arrow is not damaged Reposition the base plate and fasten it in place by tightening the t...

Page 13: ...pside down so that it rests on the platform cover Remove the screws from the base plate see arrows using the Allen wrench supplied and then remove the base plate Unscrew the four screws of the angular...

Page 14: ...r plate and remove the cover plate from the support arm base Mount the display unit on the support arm using the two retainers and the four screws supplied Route the cable through the cable conduit in...

Page 15: ...the mounting frame of the support arm to the bottom of the scale using the two M4x8 screws supplied Refasten the rectangular cover plate to the support arm using the four screws Reposition the base p...

Page 16: ...ant Note Only use original Sartorius AC adapters identified by the Sartorius la bel Use of AC adapters supplied by other manufacturers even if these adapters have a registered approval rating from a n...

Page 17: ...unded for Operation The optional interface see Interfacing Devices on page 26 in addi tion is also electrically connected to the scale housing ground Connecting Electronic Devices Peripherals Unplug t...

Page 18: ...does not require warmup POWER OFF The scale was disconnected from line power power failure or outage reconnection to line power after scale was unplugged CAL The calibration function has been activate...

Page 19: ...es you a variety of other menu definable international weight unit options Select the weight unit you need from the table of the menu options for the scale operating program Set the appropriate code a...

Page 20: ...f it has not been operated after another 2 minutes Calibration This is only possible with an accurate calibration weight 10 kg See Accessories Unload the scale and press the CAL key 7 for at least 3 s...

Page 21: ...program With the display turned off STANDBY state hold down the tare con trol 6 and briefly press the ON OFF key 8 Upon completion of the automatic self test release the tare control as soon as CH5 is...

Page 22: ...numbers for the word last digit will cycle in the display Repeat the procedure to enter the last digit of the code The o symbol that appears indicates the actual setting To change any settings C mode...

Page 23: ...e Display Format c 5 2 2 Singapore taels tl C 1 3 1 Last decimal ON c 5 2 3 Taiwan taels tl C 1 3 2 Last decimal OFF c 5 2 4 Grains gr C 1 3 3 Last decimal at stability c 5 2 5 Pennyweights dwt C 1 3...

Page 24: ...s not have a stable weight Set up scale in another area Access the menu to adapt scale to the particular type of weighing environment The weight display shows CE A zero readout was not indicated in th...

Page 25: ...re allowed to perform maintenance and repair work on the equipment Safe Operation of the scale with the AC adapter power supply is no longer ensured when there is visible damage to the AC adapter powe...

Page 26: ...ormation bulletin no 15 88 YRB 02 Z Adapter for hook up to a 12 Volt battery YCC 01 0007 Roller track conveyer plate YRT 01 IB 0001 Remote display can be connected to the interface port LCD reflective...

Page 27: ...C 40 o C 32 o F 104 o F Moisture proof rating according to DIN 40040 Class F non condensing Dust and water protection rating in compliance with DIN 40050 IEC 529 IP 65 scale NEMA 4 IP 20 AC adapter S...

Page 28: ...Copyright by Sartorius AG G ttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartori...

Reviews: