27
Specifications
Model
IB 16000 S
IB 31000 P
IB 31
Capacity kg
16,1
4/8/16/31
31
Readability
g 0,1
0,1/0,2/0,5/1
1
Maximum overload capacity
kg
75
75
75
Tare range (by subtraction)
kg
16,1
31
31
Standard deviation
g
≤
±0,05
≤
±0,1/0,1/0,25/1
≤
±0,5
Max. linearity deviation
g
≤
±0,2
≤
±0,2/0,2/0,5/1
≤
±1
Stabilization time (typical)
s
2
Adaption to ambient conditions
and application requirements
by selection of one of four digital filter levels
Display update rate
s
0,1 – 0,8 (selectable)
Stability range
d
0,25 … 64 (selectable)
Ambient temperature range
0 .
o
C.. +40
o
C (32
o
F … 104
o
F)
Moisture-proof rating according
to DIN 40040
Class F, non-condensing
Dust and water protection rating in
compliance with DIN 40050/IEC 529
IP 65 (scale) = NEMA 4
IP 20 (AC adapter)
Sensitivity drift 10 ... +30
o
C
≤
±2,5
∙
10
-6
≤
±4
∙
10
-6
≤
±10
∙
10
-6
Platform size
mm 417 x 307 (16.4 x 12.1 in.)
Scale base (W x D x H)
mm 417 x 307 x 113
(16.4 x 12.1 x 4.4 in.)
(without display unit)
Net weight of the scale
kg
15 (33 lbs.)
Net weight of the display unit
kg
1 2.2 lbs.)
Power requirements (v
frequency: 50 – 60 Hz)
115 or 230 V depending on the AC adapter used
Allowable voltage fluctuation
-20% ... +15%
Power consumption
VA 9 (typical)
Interface (optional)
RS 232 C/V24 – 28, RS 423/V10; 7-bit;
parity: even, mark, odd, space;