78 Cubis
®
MCA Operating Instructions
Technical Data
15.7.2 Models MCA6202S | MCA6202P | MCA4202S | MCA2202S
MCA6202S
MCA6202P
MCA4202S
MCA2202S
Unit
Value
Value
Value
Value
Scale interval (d)
mg
10
10 | 20 | 50
10
10
Maximum capacity (Max)
g
6200
1500 | 3000 |
6200
4200
2200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg
7
7
7
7
Standard deviation of the load values,
typical value
mg
4
4
4
4
Repeatability at approx . maximum capacity
Standard deviation of the load values,
tolerance
mg
7
40
7
7
Standard deviation of the load values,
typical value
mg
4
15
4
4
Linearity deviation
Tolerance
mg
20
50
20
20
Typical value
mg
6
20
6
6
Deviation when load is off-center, positions
according to OIML R76
Test weight
g
2000
2000
2000
1000
Tolerance
mg
20
30
30
20
Typical value
mg
10
30
10
10
Sensitivity drift b10°C and +30°C
ppm/K
2
2
2
2
Tare maximum capacity: Less than 100% of
maximum capacity
Accuracy class according to Directive 2014/31/EU
II
II
II
II
Verification scale interval (e) according to
Directive 2014/31/EU
g
0 .1
0 .1
0 .1
0 .1
Minimum load (Min) according to Directive
2014/31/EU
g
0 .5
0 .5
0 .5
0 .5
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap . 41
Optimum minimum initial weighing
g
8 .2
8 .2
8 .2
8 .2
Typical minimum initial weighing
g
8 .2
8 .2
8 .2
8 .2
Typical stabilization time
s
1
1
1
0 .8
Typical measurement time
s
1 .5
1 .5
1
1
Summary of Contents for Cubis MCA10202S Series