Cubis
®
MCA Operating Instructions
67
Cleaning and Maintenance
9 Cleaning and Maintenance
9.1 Preparing the Device
Procedure
t
Turn the device off .
t
Disconnect the device from the power supply . To do so, disconnect the power
supply cable from the wall outlet .
t
Remove the weighing pan and all associated components from the weighing
module, e .g . shield ring, pan support .
9.2 Cleaning the Device
NOTICE
Corrosion or damage to the device due to unsuitable cleaning agents!
t
Do
not
use corrosive, chloride-containing, or aggressive cleaning agents .
t
Do
not
use cleaning agents that contain abrasive ingredients, e .g . scouring
agents, steel wool .
t
Do
not
use solvent-based cleaning agents .
t
Check whether the cleaning agents used are compliant materials (see
Chapter “15 .4 Materials”, page 76) .
t
Make sure that cleaning materials, such as cloths, are only slightly damp .
Procedure
t
NOTICE
Malfunction or damage to the device due to the ingress of moisture or
dust!
t
Remove dust and powdery sample residue with a brush or hand-held vacuum
cleaner .
t
NOTICE
Corrosion or damage to the weighing pan due to unsuitable cleaning
agents
t
Wipe down the weighing pan and the shield ring with a cleaning agent and
a cloth . The cleaning agent must be suitable for stainless steel and titanium .
t
Wipe the associated components of the weighing pan with a brush or a damp
cleaning cloth, e .g . shield ring, pan support .
t
Wipe the device housing with a damp cloth . For more severe contamination,
use a mild soap solution .
Tip
We recommend cleaning the weighing pan on a regular basis, e .g . weekly . Do
not
allow deposits to form on the weighing pan .
Protective oil may be applied to the weighing pan for additional protection .
The protective oil must be suitable for stainless steel and titanium .
Summary of Contents for Cubis MCA10202S Series
Page 2: ......
Page 85: ...Cubis MCA Operating Instructions 85 Conformity Documents...