![Sartorius BP 110-0CE Installation And Operating Instructions Manual Download Page 71](http://html1.mh-extra.com/html/sartorius/bp-110-0ce/bp-110-0ce_installation-and-operating-instructions-manual_1199652071.webp)
Model
BP 610
BP 410
Weighing capacity
g
610
410
Readability
g
0.01
0.01
Tare range (by subtraction)
g
–610
–410
Reproducibility (standard deviation)*
g
≤
±0.01
≤
±0.01
Linearity
g
≤
±0.01
≤
±0.01
Response time (average)
s
≤
1
Allowable ambient operating temperature
°C
0 ... +40
Operating temperature range
°C
+10 ... +30
Sensitivity drift 10 °C ... +30 °C
/°C
≤
±5 ·10
–6
Pan size
mm
Ø 115
Dimensions (WxDxH)
mm
204x297x80
Net weight, approx.
kg
1.9
External calibration weight
(of at least accuracy class...)
g
500 (F2)
200 (F2)
Power requirements
V~
Via AC adapter: STNG 6, 230 VAC or 115 VAC,
–20% ... +15% (IP 20 protection)
Frequency
Hz
48–63
Power consumption (average)
VA
16 maximum; 8 average
Hours of operation with fully charged
YRB 05 Z external battery pack, approx.
h
45
Built-in interface
RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
AC adapter, varies acc. to country
x
x
Level indicator
x
x
Hanger for below-balance weighing
x
x
* = standard deviation of the reproducibility acc. to DIN 8120, Part 3
71