![Sartorius BP 110-0CE Installation And Operating Instructions Manual Download Page 45](http://html1.mh-extra.com/html/sartorius/bp-110-0ce/bp-110-0ce_installation-and-operating-instructions-manual_1199652045.webp)
Counting
Counting
Code 2 1 4
Symbol displayed:
P
The counting program allows automatic conversion of weights into piece counts
based on a reference sample weight. A weight readout is stored as a reference
sample quantity (factory setting: 10 pcs = pieces). When you turn on the balance,
the reference sample quantity will be displayed as “rEF 10 pcs” before you enter
the piece count.
Displaying the Reference Sample Quantity
You can have the reference sample quantity (pcs) setting displayed by pressing
the
v
key when the balance is unloaded.
Changing the Reference Sample Quantity
You can change the reference sample quantity in cycles. Choose from the
following settings: 5,10, 20, 50 and 100.
To activate the
Hold down
v
(8)
for more than 2 seconds until
change function:
“rEF…pcs” appears on the display
To change the setting:
Briefly press
v
To store the quantity:
Hold down
v
for more than 2 seconds to store
this quantity permanently (in the non-volatile memory)
after you turn off the power
This setting is
not
canceled by the reset code 9 – – 1°!
Storage Parameter for the Reference Sample Weight
The reference weight is stored…
Code
with full accuracy according to the internal resolution
3 5 1*
according to the display accuracy
3 5 2
When you exit the reference storage mode, “Err 22” may briefly be displayed.
This means that the new reference value has been stored.
Toggling between the Piece Count (pcs) and Weight (g)
After placing the sample on the weighing pan, you can toggle between the piece
count and the respective weight readout by pressing the
v
key.
*= factory setting
45