Calibration/Adjustment
During calibration
1
), the span of the balance is adjusted to the changes in
ambient conditions.
You must adjust or calibrate your new balance at the place of installation after
each warmup period and before the first measurement. You must also re-adjust
or recalibrate your balance each time you set it up in a different area or when
ambient conditions change (especially the temperature). Verified balances
approved for use as legal measuring instruments must be adjusted/calibrated
at least once a day.
The balance offers you various adjustment and calibration functions. You can
select these functions by setting the appropriate menu codes. For more information,
refer to “Balance Operating Menu.”
You can interrupt any adjustment or calibration function by pressing
c
(10)
.
Internal Calibration for Balances with a Built-in
Calibration Weight
Menu code selection: 1 9 3**
When the display shows a zero readout, press the
q
key
(9)
to activate the calibration function.
1
) “Calibration” technically means to determine
the difference between the balance readout and
the actual weight on the pan to determine the
accuracy. Adjustment means to bring a balance
into the state of accuracy required for its use.
Therefore, “calibration,” as used in this manual,
actually means “adjustment.”
*
*
= including the Signatories of the Agreement on
the European Economic Area
** = factory setting for standard balances with
a readability of
≤
0.1mg and for verified
balances approved for use as legal measuring
instruments in Europe.
Using Verified Balances as Legal Measuring Instruments in the EU*:
Before using your balance as a legal measuring instrument, you must carry
out an “Internal Calibration” operation at the place of installation after the
warmup period.