Leveling the Balance Using the Level Indicator
At the place of installation, level the balance using
the leveling feet
(5)
so that the air bubble is centered
within the circle of the level indicator
(17)
.
For balances with a rectangular pan and a weighing
capacity of <10 kg:
Retract the two auxiliary feet located at the back
of the balance.
Use the level indicator as a guide to level the balance
as follows:
To lift the balance, extend the front leveling feet
(turn clockwise).
To lower the balance, retract the front leveling feet
(turn counterclockwise).
For balances with a rectangular pan and a weighing
capacity of <10 kg:
After rectracting the rear auxiliary feet, extend them
until they touch the surface on which the balance rests.
19