Start-up
5.8 Blade handling
To insert a blade into the device, proceed as follows:
Insert the blade by handling it with fingers avoiding touching the centre.
−
Ensure the blade is completely pushed to the end of the blade holder.
Pay attention! The blade must be pushed completely to the end to avoid
risk of burning and damaging the device .
Follow the instruction displayed on the screen for the steps until the weld is
completed and the tubing removed.
1. Take the blade remover tool in your hand.
2. Push the blade remover tool over the blade.
3. Pull it back together with blade.