Getting Started
46 Arium
®
Smart Station Operating Instructions
6.1.11 Securing Protective Tubing
Procedure
t
Secure the protective tubing (1) in the clip (2) by applying pressure.
y
The clip closes, engages, and secures the protective tubing.
6.2 Connecting the Power Supply Unit
WARNING
Danger of electric shock due to incorrect handling of power cables!
Severe injuries caused by using defective power supply cables.
t
Check the power supply cable for damage, e.g., cracks in the insulation,
before use.
t
Check whether the country-specific power plug matches the power
supplies at the installation site.
t
If required: Replace the country-specific power plug adapter.
t
Only use standardized cables with a protective grounding conductor
during operation.
t
Never
disconnect the device from the protective grounding conductor.
t
Carry out the power supply according to the regulations applicable in
the relevant country.
t
Never
connect the power cable to the mains socket separately from the
device.
t
Make sure that the power plug or a different, suitable disconnecting
device for the power is easily accessible in the event of danger.
1
2