Operating Design
Arium
®
Smart Station Operating Instructions
29
Parameter
Setting Values
Explanation
Reminders
Displays the data for the next required replacement of all
consumables in the entire system.
Service
information
Service contact
Displays the responsible contact at Sartorius Service.
Next maintenance
If a service interval is activated: Displays the date for the next
scheduled maintenance (depending on the setting on the
Arium
®
water treatment device).
4.11.2 Parameters in the “Settings | Display Temperature” Menu
Parameter
Setting Values
Explanation
Display
temperature
On
−
Displays the water temperature in the operating display.
−
Can only be selected if the water quality on the Arium
®
water treatment device is displayed compensated to 25°C.
Off*
Deactivates the water temperature display.
* Default setting
4.11.3 Parameters in the “Settings | Final Filter Reminder” Menu
Parameter
Setting Values
Explanation
Reminder
Activated*
Activates the reminder for replacing the final filter.
Deactivated
Deactivates the reminder for replacing the final filter.
Filter type
Sterile filter*
Is used to select the final filter type.
Ultrafilter
Interval [months]
Entry of the interval in
months
Defines the time interval for the reminder. Is set to one month
by default.
* Default setting
4.11.4 Parameters in the “Settings | Flow Rate” Menu
Parameter
Setting Values
Explanation
Flow rate
Reset the volume of the
water dispensed
For “Pure” version only: Sets the counter back to zero.
Flow sensor adjustment
Starts the adjustment of the flow sensor.
Flow speed adjustment
For “Ultrapure” version only: Starts the adjustment of the flow
speed.