Insert and remove blades
temperature sensor
Caution:
Caution:
Caution:
Caution:
The blade has sharp edges.
Hold the blades from the sides
only.
5.3
5.3
5.3
5.3
Insert hoses
Insert hoses
Insert hoses
Insert hoses
1. Open the cover
2. Clean the outside of the tubing area to be connected using a suitable
disinfectant.
3. Place the tubing firmly in the holders. The tubing ends must protrude
beyond each holder.
4. Close both tube holders.
5.4
5.4
5.4
5.4
Insert and remove blades
Insert and remove blades
Insert and remove blades
Insert and remove blades
Insert blades
Remove a new cutting blade from its packaging. Hold the middle of the blade
between two fingers. Insert the blade into the blade holder and check that the
blade locks in correct position and the point (shown in the left picture) shows to
the temperature sensor. Insert blades into these clips
The point on the blade must
show to the temperature sensor.
– Edge with curved corners should
face up
– Please make sure that the dot
matches to the temperatur sensor
To remove a blade:
To remove a blade:
To remove a blade:
To remove a blade:
First open the cover and then attach the blade remover. You remove the blade by
pressing to the right and pulling.
Biowelder until the serialnumber RW1299 have a blade holder with runners.
Biowelder from the serial number RW1300 are equipped with balls, which the
blade pinches, it does not change the overall approach to the welding.