- 26 -
16. Actions and settings after reset
Actions and settings when a power is supplied (Power on reset)
Alarm Off
Compressor Off
Remote alarm
Off
Ring Back
30 minutes
Door alarm delay time
2 minutes
Timer Reset
2H, 8H timer
Reset (0)
Counting time during compressor L is inactive
Reset
Setting data
Read by non-volatile memory
Momentary power interruption:
When a chamber temperature is lower than set temp10
䉝
, it is diagnosed that unit is
in momentary power interruption.
Actions and settings after unit retrieves from power failure alarm:
Alarm Off
Compressor Off
Remote alarm
On
Door alarm delay time
2 minutes
Timer Reset
2H, 8H timer
Reset (0)
Counting time during compressor L is inactive
Reset
Setting data
Read by non-volatile memory
17. Lamp and display
Lamp
specifications:
<Control PCB>
Lamp
Color
Lamp status
Unit status
On
Not in alarm
DP1 Orange
Off
High/low temp. alarm, sensor failure, power failure
On
Compressor L is active
DP2 Green
Off
Compressor L is inactive
On
Capillary heater is active
DP3 Red
Off
Capillary heater is inactive
DP4 (Unused)
On
Compressor H is active
DP6 Green
Off
Compressor H is inactive
<Display PCB>
Lamp
Color
Lamp status
Unit status
DP51 Red
Flash Alarming
DP52
Green
On
Status 1 or 2 or 3 is occurred
DP53
Orange
On
Filter alarm is occurred
DP54
Red
On
Door is ajar
On
Notice of battery life
DP55 Orange
Flash
Notice of fan motor life
DP56 (Unused)