- 14 -
5. Error
codes
E01: Temp. sensor is open circuited
E02: Temp. sensor is short circuited
E03: Cascade sensor is open circuited
E04: Cascade sensor is short circuited
E05: Filter sensor is open circuited
E06: Filter sensor is short circuited
E07: AT sensor is open circuited
E08: AT sensor is short circuited
E09: Battery switch is in off position
E10: Compressor temperature is abnormal
(1) Temp. sensor
Open circuit (E01): When a temp. sensor temperature is higher than 50
䉝
, E01 and 50
䉝
are
displayed alternately and audible alarm sounds intermittently and remote
alarm contact is active.
Compressor is forcible to turn on.
Press BUZZER key to silence audible alarm.
Short circuit (E02): When a temp. sensor temperature is lower than -170
䉝
, E02 and -170
䉝
~
-180
䉝
are displayed alternately and audible alarm sounds intermittently and
remote alarm contact is active.
Compressor is forcible to turn on.
Press BUZZER key to silence audible alarm.
(2) Cascade sensor
Open circuit (E03): When a temperature in cascade sensor is lower than -65
䉝
, E03 and
chamber temperature are displayed alternately, audible alarm sounds
intermittently and remote alarm contact is active.
Both High and Low side compressors are turned off.
Press BUZZER key to silence audible alarm.
Short circuit (E04): When a temperature in cascade sensor is higher than 60
䉝
, E04 and
chamber temperature are displayed alternately, audible alarm sounds
intermittently and remote alarm contact is active.
Both High and Low side compressor are turned off.
Press BUZZER key to silence audible alarm.
(3) Filter sensor
Open circuit (E05): When a filter sensor temperature is lower than -60
䉝
, E05 and chamber
temperature are displayed alternately, audible alarm sounds intermittently
and remote alarm contact is active.
High side compressor is forcible to turn off.
Press BUZZER key to silence audible alarm.
Short circuit (E06): When a temperature in filter sensor is higher than 130
䉝
, E06 and chamber
temperature are displayed alternately, audible alarm sounds intermittently
and remote alarm contact is active.
Press BUZZER key to silence audible alarm.
(4) AT sensor
Open circuit (E07): When a temperature in AT sensor is lower than -60
䉝
, E07 and chamber
temperature are displayed alternately, audible alarm sounds intermittently
and remote alarm contact is active.
Press BUZZER key to silence audible alarm.
Short circuit(E08): When a temperature in AT sensor is higher than 60
䉝
, E08 and chamber
temperature are displayed alternately, audible alarm sounds intermittently
and remote alarm contact is active.
Press BUZZER key to silence audible alarm.
(5) Battery SW is in
off position (E09):
If you press ALARM TEST key when battery switch is in off position or
battery is unconnected, E09 is displayed.
Summary of Contents for MDF-U55V