- 18 -
Diagnosed running rate
=(-(Set temp.)X0.9)+((ATX0.9-4.5
䉝
))-((Set temp.+85
䉝
) / 10))
You can know running status by the calculation for diagnosed running rate.
Diagnosis is done by comparing running rate for 8 hours operation after 2 hours
elapses since unit started cycle operation, with diagnosed running rate.
(Diagnosed running rate – running rate) < 0 … Normal (DP52 goes off)
(Diagnosed running rate – running rate)
䍻
0 … Overload operation (DP52 illuminates)
222 = Running rate cannot be obtained by calculation. (DP52 goes off)
Diagnosis is done in every 8 hours.
10. Function
mode
F00
Display of chamber temperature (Unused)
F01
Setting of high temperature alarm
F02
Setting of low temperature alarm
F03
Display of battery accumulation time
F04
Setting of door alarm delay time
F05
Setting of compressor delay time
F06
Setting of service code (Code: 384), Reset of accumulation time
F07 *
Temperature sensor Zero Adjustment
F08 *
Cascade sensor Zero Adjustment
F09 *
Compressor continuous running mode … Factory test mode (Unused)
F10 *
Program running mode … Factory test mode (Unused)
F11 *
PCB test mode … Factory test mode (Unused)
F12 *
Display of temperature in temp. sensor
F13 *
Display of temperature in cascade sensor
F14 *
Display of temperature in filter sensor
F15 *
Display of temperature in AT sensor
F16 *
Display of chamber temperature (Unused)
F17 *
Display of model code and initialization of non-volatile memory
F18 *
Forcible operation for capillary heater
F19 *
Setting of capillary heater ON time … Factory use
F20 *
Setting of diagnosed value of overload running ratio … Factory use
F21
Setting of communication ID
F22
Setting of communication mode
F23 *
Display of power supply voltage
F24
Remote alarm terminal output
F25
Setting of Ring Back time
F26 *
Display of actual operation ratio
F27 *
Display of diagnosed value for overload running ratio
F28 *
Display of delay time of permission to measure running rate (2 hrs timer)
F29 *
Display of delay time of permission to measure running rate (8 hrs timer)
F30 *
Display of ROM version
F31 *
Setting of filter alarm
F32
Display of condensing fan motor accumulation time
F33~F43
Unused
F44 *
Adjustment of power supply voltage display
F45~F49
Unused
F50
Setting of alarm delay time
F51~F59
Unused
Input service code ‘384’ in F06 prior to use function codes which are marked with *.
To cancel service code, input “000” in F06 or turn the power off.
Setting
procedure:
In chamber temperature display, press key for 5 seconds to display “F00”.
Input Function code by pressing key and key.