5 4
―
―
5 3
―
DC/AC Voltage (DC/AC V) for dual display
Range
Accuracy
50Hz
〜
60Hz
60.00mV, 600.0mV
± (0.7% rdg + 6dgt)
9.999V, 99.99V, 999.9
40Hz
〜
1kHz
60.00mV, 600.0mV
± (1.0% rdg + 6dgt)
9.999V, 99.99V, 999.9V
± (2.2% rdg + 6dgt)
1kHz
〜
20kHz
60.00mV, 600.0mV
± (2.2% rdg + 6dgt)
9.999V
1)
, 99.99V
± 3dB
999.9V
Un specified
Input impedance: 10M
Ω
, 80pF nominal
(130pF nominal for 600.0mV&60.00mV range)
1) 3kHz
〜
15kHz
DC current (DCA)
Range
Accuracy
Input resistance* *
600.0
µ
A
±
(
0.2% rdg + 4dgt
)
Approx. 83
Ω
6000
µ
A
60.00mA
Approx. 1
Ω
600.0mA
6.000A
Approx. 0.005
Ω
10.00A*
* > 6A: Cool down more than 3 minutes after measuring 1 minute.
< 6A Continuable
* * Fusing resistor not included
AC current (ACA), DC/AC current (DC/AC A)
Range
Accuracy
Input resistance* *
DC, 50Hz to 60Hz
600.0
µ
A
±
(
0.6% rdg + 3dgt
)
Approx. 83
Ω
6000
µ
A
60.00mA
Approx. 1
Ω
600.0mA
±
(
1.0% rdg + 3dgt
)
6.000A
10.00A*
±
(
0.8% rdg + 6dgt
)
Approx. 0.05
Ω
40Hz to 1kHz
600.0
µ
A
6000
µ
A
±
(
0.8% rdg + 4dgt
)
Approx. 83
Ω
60.00mA
Approx. 1
Ω
600.0mA
±
(
1.0% rdg + 4dgt
)
6.000A
10.00A*
±
(
0.8% rdg + 6dgt
)
Approx. 0.005
Ω
* > 6A: Cool down more than 3 minutes after measuring 1 minute.
< 6A Continuable
**Fusing resistor not included