2 0
―
―
1 9
―
Note:
・
The display style of [
/
] does not show the bar graph.
5-4
「
」
(Max. rated input voltage: 10V dc/ac)
・
DC voltage (m
) measurement
・
DC Voltage (m
)/AC Voltage(m
) simultaneous measurement
・
Logic-level frequency (
) measurement
・
Duty cycle (
D%) Measurement
1. Do not apply any input signal exceeding the max. rated input
voltage.
2. Do not switch the function selector while measuring.
3. Keep your fingers behind the finger guards of the test leads while
measurement.
WARNING
1) What to measure
・
m
(DC voltage): DC circuit voltage lower than 600mV
・
m
/m
(DC voltage component/AC voltage component)
・
(Logic level frequency): 3V, 5V logic circuit frequency
・
D%(Duty cycle): Logic level signal duty cycle (Square wave)
2) Measuring ranges
・
m
, m
/m
: 60.00mV and 600.0mV
・
: Auto ranging, 5.000Hz to 1.000MHz (Square wave)
・
D%: 0.00% to 100.0% (Square wave 5Hz to 10kHz)
3) Measuring procedure
①
Connect the red plug of the test lead to the VHz measuring
terminal and the black one to the COM terminal.
②
Set the function selector to
.
③
Press the SELECT button to select a function you want to
perform.
④
Apply the test pins (Red and Black) to the object to measure.
⑤
Read the display.