Inductance: Lp, Ls (When D < 0.1)
Range Resolution 100/120 Hz
1 kHz
10 kHz
100 kHz
20 μH 0.001 μH
-
-
-
1.0 % + 10 d
200 μH 0.01 μH
-
-
0.8 % + 10 d 1.0 % + 10 d
2000 μH 0.1 μH
-
0.8 % + 10 d 0.5 % + 5 d 0.8 % + 10 d
20 mH 0.001 mH 0.8 % + 10 d 0.3 % + 3 d 0.5 % + 5 d 0.8 % + 10 d
200 mH 0.01 mH 0.5 % + 5 d 0.5 % + 5 d 0.8 % + 10 d
-
2000 mH 0.1 mH
0.3 % + 3 d 0.5 % + 5 d 0.8 % + 10 d
-
20 H
0.001 H 0.3 % + 3 d 0.3 % + 3 d 5.0 % + 20 d
-
200 H
0.01 H
0.5 % + 5 d 0.8 % + 10 d
-
-
2000 H
0.1 H
0.8 % + 10 d 3.0 % + 20 d
-
-
20 kH 0.001 kH 5.0 % + 20 d
-
-
-
• Accuracy when D
≧
0.1: Ae×
Ae means an accuracy of rdg in the main display.
Accuracy of Θ θe: θe=±(180/π) × Ae (deg)
Ae means an accuracy of rdg in the main display.
• In the capacitance measurement
Range
100/120 Hz
1 kHz
10 kHz
100 kHz
200 pF
-
-
±0.46
±1.15
2000 pF
±0.29
±0.29
±0.29
±0.40
20 nF
±0.29
±0.17
±0.29
±0.17
200 nF
±0.17
±0.29
±0.29
±0.40
2000 nF
±0.17
±0.17
±0.29
±2.01
20 μF
±0.17
±0.17
±0.57
-
200 μF
±0.17
±0.46
±2.01
-
2000 μF
±1.14
±0.86
-
-
20 mF
±0.86
-
-
-
• In the inductance measurement
Range
100/120 Hz
1 kHz
10 kHz
100 kHz
20 μH
-
-
-
±0.57
200 μH
-
-
±0.46
±0.57
2000 μH
-
±0.46
±0.29
±0.46
20 mH
±0.46
±0.17
±0.29
±0.46
200 mH
±0.29
±0.29
±0.46
-
2000 mH
±0.17
±0.29
±0.46
-
20 H
±0.17
±0.17
±2.87
-
200 H
±0.29
±0.46
-
-
2000 H
±0.46
±1.72
-
-
20 kH
±2.87
-
-
-
― 40 ―
― 39 ―