5-4-2 Capacitance (C) measurement
1) Measuring ranges
C: 200.00 pF
~
20.00 mF (auto-range)
2) Measuring procedure
①
Press the
L/C/R AUTO/MANU
button to select Cs or Cp function.
②
Press the
TEST FREQ
button to select a measuring frequency.
③
Press the
S/P
button to select Cs or Cp.
④
Connect a DUT to the measuring terminals.
⑤
Press the
D/Q/θ/ESR/Rp
button to select a sub parameter.
The sub parameter in the Cs function includes D, Q, ESR, and θ.
The sub parameter in the Cp function includes D, Q, Rp, and θ.
⑥
Read the display.
5
-4-1 Inductance (L) measurement
1) Measuring ranges
L: 20.000 μH
~
2000 H (auto-range)
2) Measuring procedure
①
Press the
L/C/R AUTO/MANU
button to select Ls or Lp function.
②
Press the
TEST FREQ
button to select a measuring frequency.
③
Press the
S/P
button to select Ls or Lp.
④
Connect a DUT to the measuring terminals.
⑤
Press the
D/Q/θ/ESR/Rp
button to select a sub parameter.
The sub parameter in the Ls function includes Q, ESR, θ, and D.
The sub parameter in the Lp function includes Q, Rp, θ, and D.
⑥
Read the display.
Example